JF 17.07.13
Service Manual BB 150
Index
Name
Date
Document No.
Page
File
500789823_05_CN
50 079 823
1 / 1
5. INITIAL SETUP
Instruction
Input / key(s) / comments
Display / comment /status
Open all doors.
Remove shipping restraints
and fixed equipment from
sample chamber.
Clean the unit.
Water reservoir
Add water.
Do not exceed max. level,
use only distilled sterile water.
Switching on the unit
Set the main power switch to
its "I" position.
All indicators on the control
panel go on for approx. 15
sec. (8-digit test).
Software versions and
parameter set are shown.
Setpoint specification
Temperature setpoint adjustment
Display temperature setpoint.
Press
°C
Current setpoint is displayed,
digit to the right of decimal
flashes.
Enter new temperature
setpoint.
Press
°C
&
Increase the setpoint.
Press
°C
&
Decrease the setpoint.
Save NEW temperature
setpoint.
Release
°C
Sample chamber temperature
is displayed.
CO
2
setpoint adjustment
Display CO
2
setpoint.
Press
Current setpoint is displayed,
digit to the right of decimal
flashes.
Enter new CO
2
setpoint.
Press
&
Increase the setpoint.
Press
&
Decrease the setpoint.
Save NEW CO
2
setpoint.
Release
CO
2
concentration in the
sample chamber is displayed.