Safety precautions
8
GenPure
Thermo Scientific
• If the system is to be at a rest for a longer time (e.g. long, holidays) switch the system off (unplug
the mains plug) and shut off the feedwater line.
Allowing the system to run with the water feed line closed would result in damage to the pump.
The manufacturer does not accept liability for such damage.
• Unplug the system from the power outlet for all Maintenance works on the system.
• The system must be subjected to rinsing and possibly also disinfection after longer rest periods.
Follow the directions given in the section
“Rinsing the ultrafilter”
on
page 71
.
• The surface or wall where the system is to be installed must have adequate load-bearing capacity
(check the capacity and stability of the wall). The weight of the ultrapure water system is given
under
“Dimension and weight GenPure System”
on
page 20
.
• The surface on which the unit is installed must be level and stable - not to exceed a maximum of
2% deviation from evenness is recommended.
• When installing the ultrapure water system, always ensure that there is adequate space all around
the unit (approx 30cm / 11.81 inch) to ensure that ease of use or replacement of materials (e.g.
filter change, connection) is possible at all times.
• Visually inspect the system at regular intervals. Clean up any water or spills found around the
system immediately
.
Never look directly into a switched-on UV-lamp, as UV-light endangers eyesight!
To avoid the risk of pinching, crushing cutting or electrical shock, never perform
maintenance on the unit without its protective housing, or while it is in operation.
Maintenance work on the system may only be performed by trained, authorized
specialists.
• Wear safety gloves when working with cleaning solution.
• If your skin should come into contact with a chlorine product, rinse it immediately with ample,
fresh water.
• The unit or system components, may heat up as a result of a defect. It is recommended to always
wear appropriate safety gloves to prevent skin damage or burns.
• Wear safety gloves when changing the UV-lamp, in order to prevent that your skin comes in
contact with the UV-lamp glass.
• Wear safety goggles when working with cleaning solution.
• If your eyes come into contact with a chlorine product, rinse them immediately with ample,
fresh water and immediately contact a physician.
Содержание Barnstead GenPure xCAD Plus System
Страница 10: ...Contents 4 GenPure Thermo Scientific ...
Страница 16: ...2 Safety precautions 10 GenPure Thermo Scientific ...
Страница 22: ...3 Extent of delivery Available GenPure xCAD Plus versions 16 GenPure Thermo Scientific ...
Страница 24: ...4 Intended Use 18 GenPure Thermo Scientific ...
Страница 30: ...5 Technical specifications 24 GenPure Thermo Scientific ...
Страница 32: ...6 The installation area 26 GenPure Thermo Scientific ...
Страница 58: ...8 52 GenPure Thermo Scientific ...
Страница 60: ...9 How the system functions 54 GenPure Thermo Scientific ...
Страница 104: ...13 Maintenance Change and autoclave the Final filter 98 GenPure Thermo Scientific ...
Страница 106: ...14 Waste disposal 100 GenPure Thermo Scientific ...
Страница 110: ...15 Trouble shooting 104 GenPure Thermo Scientific ...
Страница 111: ...Thermo Scientific GenPure 105 16 Replacement parts GenPure 1 2 9 3 12 6 15 8 11 13 4 7 5 14 10 ...
Страница 114: ...16 Replacement parts GenPure 108 GenPure Thermo Scientific ...
Страница 116: ...17 Consumable materials 110 GenPure Thermo Scientific ...
Страница 118: ...18 Accessories 112 GenPure Thermo Scientific ...
Страница 119: ...Thermo Scientific GenPure 113 19 Terminal assignments ...
Страница 120: ...19 Terminal assignments 114 GenPure Thermo Scientific ...
Страница 122: ...20 Maintenance records 116 GenPure Thermo Scientific ...