12
System control
General information
64
GenPure
Thermo Scientific
General information
The software structure consists of five operating modes and four menus, which will be described in
more detail in the following sections. Measured values are continually shown in the display and/or in
the menus. The displayed TOC value is calculated from the difference in the ultrapure water
measuring cell and TOC-measurement measuring cell values.
When an error occurs, the display backlighting changes from green to red and the error message is
shown in clear text in the first line of the display in alternation with the operating mode message.
Operating modes
Interval operating mode after switching on
Initially press the ON/OFF button. Then the display will show at first the system version, the system
serial number and the software version number to display for 3 seconds. The system then automatically
switches to the Interval operating mode (see
“Interval operation”
on
page 65
), whereby the green
backlighting of the display is switched on and remains on until system control is switched off via the
ON/OFF-button. The “UV” text message is displayed when the UV-lamp is switched on. The “TC”
message is displayed when measured values are subject to temperature compensation. Further to these,
the measured values for ultrapure water (measuring cell LF1) and temperature are also displayed. The
displays of messages and measured values are independent of the operating mode.
The TOC value is not shown in Interval mode.
The display shows:
Green display:
operation of the
system is okay.
Red display:
There is a error in
the system.
Содержание Barnstead GenPure xCAD Plus System
Страница 10: ...Contents 4 GenPure Thermo Scientific ...
Страница 16: ...2 Safety precautions 10 GenPure Thermo Scientific ...
Страница 22: ...3 Extent of delivery Available GenPure xCAD Plus versions 16 GenPure Thermo Scientific ...
Страница 24: ...4 Intended Use 18 GenPure Thermo Scientific ...
Страница 30: ...5 Technical specifications 24 GenPure Thermo Scientific ...
Страница 32: ...6 The installation area 26 GenPure Thermo Scientific ...
Страница 58: ...8 52 GenPure Thermo Scientific ...
Страница 60: ...9 How the system functions 54 GenPure Thermo Scientific ...
Страница 104: ...13 Maintenance Change and autoclave the Final filter 98 GenPure Thermo Scientific ...
Страница 106: ...14 Waste disposal 100 GenPure Thermo Scientific ...
Страница 110: ...15 Trouble shooting 104 GenPure Thermo Scientific ...
Страница 111: ...Thermo Scientific GenPure 105 16 Replacement parts GenPure 1 2 9 3 12 6 15 8 11 13 4 7 5 14 10 ...
Страница 114: ...16 Replacement parts GenPure 108 GenPure Thermo Scientific ...
Страница 116: ...17 Consumable materials 110 GenPure Thermo Scientific ...
Страница 118: ...18 Accessories 112 GenPure Thermo Scientific ...
Страница 119: ...Thermo Scientific GenPure 113 19 Terminal assignments ...
Страница 120: ...19 Terminal assignments 114 GenPure Thermo Scientific ...
Страница 122: ...20 Maintenance records 116 GenPure Thermo Scientific ...