13
Maintenance
Change the UV-lamp
Thermo Scientific
GenPure
95
7
Put the cartridge cover back on (see under section
“Change the ultrapure cartridge”
on
page 85
),
re-open the feed water supply and switch the
system on again.
8
Push the menu button until “Enter code” is
displayed.
The Code to do this transaction please refer from
the Code table under section
“Code lock”
on
page 82
. You need a level 3 code.
a.
After entering the code push the Menu
and UV button simultaneously. The
display shows UV Menu.
b. Push the Menu button repeatedly until
new UV-lamp appears and press enter to
confirm.
c.
The system sets the operating hours
counter of the UV-lamp back and save
the new values by an automatically
calibration.
The UV-lamp must be switched on (Nonstop
mode).
The calibration process of the UV intensity can
be take between 5 min. and 2 hours.
Step
Action
Figure
Feedwater supply
a.
b.
Содержание Barnstead GenPure xCAD Plus System
Страница 10: ...Contents 4 GenPure Thermo Scientific ...
Страница 16: ...2 Safety precautions 10 GenPure Thermo Scientific ...
Страница 22: ...3 Extent of delivery Available GenPure xCAD Plus versions 16 GenPure Thermo Scientific ...
Страница 24: ...4 Intended Use 18 GenPure Thermo Scientific ...
Страница 30: ...5 Technical specifications 24 GenPure Thermo Scientific ...
Страница 32: ...6 The installation area 26 GenPure Thermo Scientific ...
Страница 58: ...8 52 GenPure Thermo Scientific ...
Страница 60: ...9 How the system functions 54 GenPure Thermo Scientific ...
Страница 104: ...13 Maintenance Change and autoclave the Final filter 98 GenPure Thermo Scientific ...
Страница 106: ...14 Waste disposal 100 GenPure Thermo Scientific ...
Страница 110: ...15 Trouble shooting 104 GenPure Thermo Scientific ...
Страница 111: ...Thermo Scientific GenPure 105 16 Replacement parts GenPure 1 2 9 3 12 6 15 8 11 13 4 7 5 14 10 ...
Страница 114: ...16 Replacement parts GenPure 108 GenPure Thermo Scientific ...
Страница 116: ...17 Consumable materials 110 GenPure Thermo Scientific ...
Страница 118: ...18 Accessories 112 GenPure Thermo Scientific ...
Страница 119: ...Thermo Scientific GenPure 113 19 Terminal assignments ...
Страница 120: ...19 Terminal assignments 114 GenPure Thermo Scientific ...
Страница 122: ...20 Maintenance records 116 GenPure Thermo Scientific ...