13
Maintenance
Change the ultrafilter
90
GenPure
Thermo Scientific
Change the ultrafilter
(applicable only for systems with UF)
Step
Action
Figure
1
Switch the GenPure xCAD Plus System off and
shut off the supply of feedwater.
2
Remove the four screws of the back panel.
Remove carefully the back panel from the system
and unscrew the yellow ground wire from the
back panel.
3
a.
Pull out the hoses 8 mm by unscrew the
two fittings (see red arrows).
b. After this procedure draw out the
ultrafilter from the mounting clamp (see
red arrow).
c.
Hold with one hand the hose and with
the other hand turn in clockwise
direction the ultrafilter to unscrew the
hose connection.
d. When you are finished with step c install
the new ultrafilter by attaching hoses and
mounting it in the clamp.
e.
When you are installing the new UF
filter the flow arrow of the filter must be
pointing to the bottom of the system.
4
Reinstall the back panel, reopen the feedwater
supply and switch on the system again.
Feedwater supply
Screws
Screws
a.
b.
c.
Hose
Ultra
Filter
White
O-ring
Fitting
Mounting
clamps
Ultra
Filter
White
O-ring
Fitting
Feedwater
supply
Содержание Barnstead GenPure xCAD Plus System
Страница 10: ...Contents 4 GenPure Thermo Scientific ...
Страница 16: ...2 Safety precautions 10 GenPure Thermo Scientific ...
Страница 22: ...3 Extent of delivery Available GenPure xCAD Plus versions 16 GenPure Thermo Scientific ...
Страница 24: ...4 Intended Use 18 GenPure Thermo Scientific ...
Страница 30: ...5 Technical specifications 24 GenPure Thermo Scientific ...
Страница 32: ...6 The installation area 26 GenPure Thermo Scientific ...
Страница 58: ...8 52 GenPure Thermo Scientific ...
Страница 60: ...9 How the system functions 54 GenPure Thermo Scientific ...
Страница 104: ...13 Maintenance Change and autoclave the Final filter 98 GenPure Thermo Scientific ...
Страница 106: ...14 Waste disposal 100 GenPure Thermo Scientific ...
Страница 110: ...15 Trouble shooting 104 GenPure Thermo Scientific ...
Страница 111: ...Thermo Scientific GenPure 105 16 Replacement parts GenPure 1 2 9 3 12 6 15 8 11 13 4 7 5 14 10 ...
Страница 114: ...16 Replacement parts GenPure 108 GenPure Thermo Scientific ...
Страница 116: ...17 Consumable materials 110 GenPure Thermo Scientific ...
Страница 118: ...18 Accessories 112 GenPure Thermo Scientific ...
Страница 119: ...Thermo Scientific GenPure 113 19 Terminal assignments ...
Страница 120: ...19 Terminal assignments 114 GenPure Thermo Scientific ...
Страница 122: ...20 Maintenance records 116 GenPure Thermo Scientific ...