How the system functions
Tap water that has been pretreated upstream by reverse osmosis, ion exchange or distillation flows
through a pressure reducer and into the ultrapure water system, where the conductivity is monitored. A
pump directs this feedwater through UV-photooxidation (only possible in UV lamp equipped systems)
and then through the ultrapure cartridge. From there the water flows through an ultrafiltration module
(only possible in UF equipped systems). Then follows a permanent definition of conductivity
measured by a special conductivity measuring cell equipped with temperature compensation. When
ultrapure water is dispensed from the system, it flows through a a end filter before reaching the point of
use. During Interval operation, the water in the system is circulated in an internal circuit at regular
intervals.
Systems with UV-TOC, UV-TOC/UF
Tap water that has been pretreated upstream by reverse osmosis, ion exchange or distillation passes
through a pressure reducer and into the ultrapure water system, where the conductivity is monitored. A
pump directs this feedwater through UV-photooxidation, which follows a conductivity measurement
to determine the TOC value. Then follows an ultrapure cartridge and an ultrafiltration module (only
with UV-TOC/UF), and the conductivity is then permanently measured by a special measuring cell
(with temperature compensation). When ultrapure water is taken from the system, it flows through a
final filter before reaching the dispensing outlet. During Interval operation, the water in the system is
recirculated in an internal circuit at regular intervals.
The TOC value is calculated by taking the difference between the values measured by the measuring
cells QIA300 and QI302. The measurement range is 0 - 30 ppb. When this range is exceeded, the
number 99 is shown in the display instead of the measured value.
In Stand-by operation, “___” is shown.
System Function as applied in all GenPure systems
Содержание Barnstead GenPure xCAD Plus System
Страница 10: ...Contents 4 GenPure Thermo Scientific ...
Страница 16: ...2 Safety precautions 10 GenPure Thermo Scientific ...
Страница 22: ...3 Extent of delivery Available GenPure xCAD Plus versions 16 GenPure Thermo Scientific ...
Страница 24: ...4 Intended Use 18 GenPure Thermo Scientific ...
Страница 30: ...5 Technical specifications 24 GenPure Thermo Scientific ...
Страница 32: ...6 The installation area 26 GenPure Thermo Scientific ...
Страница 58: ...8 52 GenPure Thermo Scientific ...
Страница 60: ...9 How the system functions 54 GenPure Thermo Scientific ...
Страница 104: ...13 Maintenance Change and autoclave the Final filter 98 GenPure Thermo Scientific ...
Страница 106: ...14 Waste disposal 100 GenPure Thermo Scientific ...
Страница 110: ...15 Trouble shooting 104 GenPure Thermo Scientific ...
Страница 111: ...Thermo Scientific GenPure 105 16 Replacement parts GenPure 1 2 9 3 12 6 15 8 11 13 4 7 5 14 10 ...
Страница 114: ...16 Replacement parts GenPure 108 GenPure Thermo Scientific ...
Страница 116: ...17 Consumable materials 110 GenPure Thermo Scientific ...
Страница 118: ...18 Accessories 112 GenPure Thermo Scientific ...
Страница 119: ...Thermo Scientific GenPure 113 19 Terminal assignments ...
Страница 120: ...19 Terminal assignments 114 GenPure Thermo Scientific ...
Страница 122: ...20 Maintenance records 116 GenPure Thermo Scientific ...