The installation area
Take the following criteria into consideration when selecting the installation area:
Feedwater pressure, not below 0.1 bar (1.45 PSI) and not above 6 bar (87 PSI).
• Minimum air tempe 2 °C.
• The surface on which the system is installed must be level and stable - not to exceed a maximum of
2% deviation from evenness is recommended.
• A smooth wall is required when the system is to be wall-mounted. Check the statics of the wall. It
must have sufficient load-bearing capacity (for system weight, see
“Technical specifications”
on
page 19
).
• An atmospherically floor drain with a outer diameter of 63mm or 2.48 inch (DN 50 tube) shall be
provided. When no floor drain is available, install a water watcher to protect against water damage
(only available for EU).Otherwise the manufacturer will not accept liability for any possible water
damage.
• Free run off to drain.
• An electrical socket appropriate for the system (see
“Technical specifications”
on
page 19
).
• When installing the ultrapure water system, always ensure that there is adequate space all around
the unit (approx 30cm / 11.81 inch) to ensure that ease of use or replacement of materials (e.g.
filter change, connection) is possible at all times
• Easy access for operation and control of the system.
• Water pre treated such as DI, RO or distillation water connection with 3/4 NPT male thread and
customer supplied shutoff valve.
The operator is obliged to ensure, that the installation of the water purification unit
and its operation are carried out in compliance with all national and international
guidelines, applicable and valid for the place of installation. If necessary, measures to
protect the drinking water have to be taken by installing appropriate components.
The feedwater pressure must never exceed 6 bar. If it is higher than this, install an
additional external pressure reducer.
Unrestricted gravity flow to drain must be ensured!
Содержание Barnstead GenPure xCAD Plus System
Страница 10: ...Contents 4 GenPure Thermo Scientific ...
Страница 16: ...2 Safety precautions 10 GenPure Thermo Scientific ...
Страница 22: ...3 Extent of delivery Available GenPure xCAD Plus versions 16 GenPure Thermo Scientific ...
Страница 24: ...4 Intended Use 18 GenPure Thermo Scientific ...
Страница 30: ...5 Technical specifications 24 GenPure Thermo Scientific ...
Страница 32: ...6 The installation area 26 GenPure Thermo Scientific ...
Страница 58: ...8 52 GenPure Thermo Scientific ...
Страница 60: ...9 How the system functions 54 GenPure Thermo Scientific ...
Страница 104: ...13 Maintenance Change and autoclave the Final filter 98 GenPure Thermo Scientific ...
Страница 106: ...14 Waste disposal 100 GenPure Thermo Scientific ...
Страница 110: ...15 Trouble shooting 104 GenPure Thermo Scientific ...
Страница 111: ...Thermo Scientific GenPure 105 16 Replacement parts GenPure 1 2 9 3 12 6 15 8 11 13 4 7 5 14 10 ...
Страница 114: ...16 Replacement parts GenPure 108 GenPure Thermo Scientific ...
Страница 116: ...17 Consumable materials 110 GenPure Thermo Scientific ...
Страница 118: ...18 Accessories 112 GenPure Thermo Scientific ...
Страница 119: ...Thermo Scientific GenPure 113 19 Terminal assignments ...
Страница 120: ...19 Terminal assignments 114 GenPure Thermo Scientific ...
Страница 122: ...20 Maintenance records 116 GenPure Thermo Scientific ...