18. Remove the protective cap from the filter bell. Turn the dispense knob
to the left or right and flush 1-2 liters of water through the 0.2 micron
filter. The product water is now ready for use.
Note
For more demanding applications where low TOC water is required,
a rinse of 15-20 liters through the cartridges and 0.2 micron filter may be
necessary.
s
Caution
Never dispense water without first installing hose barb provided in
parts bag or final filter supplied with cartridge pack.
s
Nanopure™ Analytical 7-5
Thermo Scientific
Section 7
Initial Operation
System Cleaning
Procedure (cont.)
0.2 Micron
Filter
Figure 7-3.
Installing 0.2 Micron Filter