8. Press BACK to return to the (Idle) mode.
The second line of the display will show a “Below Set point” message once
every 8 to 10 seconds when the measurement is below the set point. (The
set point indication is set at the factory at 10 M
Ω
-cm.)
1. From the (Idle) mode, when the display reads, “Nanopure (Idle)
xx:xx:xx xx/xx/xx,” press the UP or DOWN arrow until the display
reads, “Reset Timers?”
2. Press ENTER.
3. Display will read, “Reset Cleaning Timer” and show on the right side
of the second line when the user will start receiving “Clean Unit”
reminders.
4. Press the UP or DOWN arrow to select “Yes.”
5. Press ENTER and the timer will be reset. This will reset the cleaning
timer for approximately six months calendar time.
Note
Volumetric dispensing is based upon total fluid volume entering the
Nanopure Analytical. Point of drawoff accuracy will be compromised if
you are dispensing water volumetrically while simultaneously drawing
water manually through the remote dispenser accessory.
s
Note
Once Volumetric or Timed Dispensing values are set, they will
remain in memory until changed by the user. To turn the dispense method
OFF, select “OFF” in the “Dispense Method” menu.
s
1. From the normal recirculating mode, when the display is showing
purity, press the UP or DOWN arrow until the display reads, “Auto
Dispense Menu?”
2. Press ENTER.
3. Display will read, “Dispense Method.” Press the UP or DOWN arrow
to select VOLUMETRIC.
4. Press ENTER. Display will read, “Unit Volume Setting: xx Liters.”
5. Press the UP or DOWN arrow to set the desired volume value. Press
ENTER.
Nanopure™ Analytical 9-5
Thermo Scientific
Section 9
User Settings
Selecting the Set
Point (continued)
Resetting the
Cleaning Timer
Setting Volumetric
Dispensing