10-4
Nanopure™ Analytical
Thermo Scientific
Section 10
Accessories
10. Display will read “Calibration (In Progress).” When the calibration is
complete the display will read “Calibration (passed/failed) for 4
seconds and then return to the”N.I.S.T. Calibrate?” display.
Note
Calibration (in progress) may take up to two minutes to complete.
s
11. Press BACK to return to the (Idle) mode.
12. Turn the main power switch to the OFF “O” position. Unplug the
unit.
13. Remove the calibration module and reconnect the cell.
14. Reattach the cover plate and close and latch the door.
15. Reconnect the power cord.
The calibration procedure is complete.
Refer to the Operating Instructions for your manual or Accudispense
volumetric remote dispenser (LT1366X11).
Remote Volumetric
Dispenser
Electronic Cal. w/
Opti. Cal. Mod. (cont.)