5-2
Nanopure™ Analytical
Thermo Scientific
Section 5
Installation
Optional accessories for this unit include a float switch, low pressure
switch, N.I.S.T. Traceable Calibration Module, standard remote dispenser,
Accudispense volumetric remote dispenser, dispense overflow cutoff float
and printer. If you purchased a float switch or low pressure switch, refer to
the Accessories section of this manual. If you purchased a standard or
Accudispense volumetric remote dispenser, refer to their respective
operator’s manuals for installation instructions.
1. Place the Nanopure Analytical on a bench top that is accessible to
pretreated water, electricity and an atmospherically vented drain.
Install the Nanopure Analytical on a wall in a convenient location that is
accessible to water, an atmospherically vented drain and electricity.
1. Locate the wall bracket packed separately from the unit.
2. Using the wall bracket as a template, locate and drill the mounting
holes in the wall. A minimum of four (customer-supplied) fasteners
will be required — two on the top and two on the bottom.
3. Attach the wall bracket to the wall using the customer-supplied
fasteners.
4. Remove the locking screws on each side of the wall bracket.
5. Pull the two locking slides on each side of the wall bracket out as far as
they will go.
6. Hang the unit on the wall bracket by sliding the mounting pins into
the wall bracket slots.
7. Push the locking slides on each side of the wall bracket in as far as they
will go.
8. Replace the locking screws.
Note
Please refer to the “General Specifications” section for clearance
requirements.
s
Other Accessories
Bench Mounting
Wall Mounting