Warning
Do not locate the Nanopure Analytical directly over equipment
that requires electrical service. Routine maintenance of this unit may
involve water spillage and subsequent electrical shock hazard if improperly
located.
s
Note
The outlet of a gravity feed storage reservoir must be above or at the
same level as the inlet of the Nanopure Analytical.
s
Caution
Wall composition, condition and construction, as well as fastener
type, must be considered when mounting this unit. The mounting surface
and fasteners selected must be capable of supporting a minimum of 150
lbs.(68kg). Inadequate support and/or fasteners may result in damage to
mounting surface and/or equipment. If you are unsure of mounting
surface composition, condition and construction or correct fasteners,
consult your building maintenance group or contractor.
s
For your convenience, the control panel can be removed from the unit and
mounted at a convenient location within 10 ft. (3.1 m.) of the unit. To
remove the control panel from the unit:
1. Turn the unit OFF and disconnect it from the power supply.
2. Slide the control panel upward about 1/4” (.64 cm) to 3/8” (.95 cm) -
(just enough to clear connector on top cover), pull forward, and
remove. (The control panel fits tightly in the unit so it may be difficult
to remove.)
Note
The removable control panel was not designed to be repeatedly
removed from the unit, therefore, you may experience difficulty when
attempting to remove it. It is therefore recommended that the control
panel be permanently mounted in a remote location or remain intact with
the unit.
s
3. Replace the control panel with the blank panel provided with the unit
to protect the electrical connector.
Note
The blank protective cover on the rear side panel display connector
should remain in place when the display is not remotely located.
Nanopure™ Analytical 5-3
Thermo Scientific
Section 5
Installation
Wall Mounting
(continued)
Install Control Panel
in a Remote Location