1300 Series A2 4-3
Thermo Scientific
Section 4
Operation
The following are the operating modes for this unit.
• OFF mode
• Window open mode
• Working mode
• Standby mode
• UV mode
Off mode
means the unit is in an idle state. The unit is turned off
(blowers are off). The sample chamber light is enabled. The unit is
connected to a power source. The power supply for the internal outlets is
enabled when the blue LED “internal power supply on” is lighted.
Window open mode
is the state in which the window is open and
positioned outside of the 250mm working position. To install or remove
accessories, the front window can be moved to the maximum open
position. For cleaning, the front window can be lowered beyond the closed
position (see Use of the Window, Section 1).
The air system blowers are switched on with the red LED “airflow not
steady” lighted. The front window is not in the work position, with the
red LED “front window not in work position” is lighted. The sample
chamber light is enabled. The power supply for the internal outlets is
enabled with the blue LED “internal power supply on” is lighted.
Working mode
is active when the front window is in the working position
(250 mm opening) and the air system is operating steadily. The green
LEDs “front window is in work position” and “airflow steady” are lighted.
The markings on the side guide rails and the lower edge of the front
window are at the same height with no audible alarm signal.
The sample chamber light and power for the internal outlets are enabled
with the yellow LED “internal power supply activated” is lighted.
The RS-232 contact is open. The display shows operating hours,
downflow velocity and exhaust air velocity.
Standby mode
is the state in which the window is closed with the blower
velocity reduced. The red LED “front window not in work position” and
the blue LED “airflow reduced” are lighted.
The sample chamber illumination is enabled. The internal power supply in
the sample chamber is enabled. The red LED “airflow not steady” and the
blue LED “internal power supply activated” are lighted.
Operating Modes
Содержание 1300 Series
Страница 25: ...Thermo Scientific Section 2 Installation 2 8 1300 Series A2 ...
Страница 31: ...Thermo Scientific Section 3 Start Up 3 6 1300 Series A2 ...
Страница 39: ...Thermo Scientific Section 4 Operation 4 8 1300 Series A2 ...
Страница 45: ...Thermo Scientific Section 5 Cleaning and Decontamination 5 6 1300 Series A2 ...
Страница 69: ...Thermo Fisher Scientific 401 Millcreek Road Marietta Ohio 45750 United States www thermofisher com ...
Страница 70: ...1300 系列 A2 II级 A2型 生物安全柜 操作手册 5991355 ...
Страница 91: ...Section 2 安装 维修阀安装 续 维修阀 维修阀插头 垫片 螺母 图2 4 阀门总成和安装 2 6 1300 Series A2 Thermo Scientific 螺纹管 底座 ...
Страница 129: ...仪器类型 部件号 序列号 维修号 地点 操作者标注 技术员 公司 记录 日期 签名 Thermo Scientific 1300 Series A2 9 1 ...