To position the window at the operating height (250 mm opening), move
it to the area of the top ‘dimple’ (Fig. 1-5) and watch for the GREEN sash
position height indicator to illuminate (Fig. 1-6). If the window is moved
from the operating height, the sash height indicator will turn RED.
To place the window in the closed or reduced mode position, move it to
the area of the lower ‘dimple’ (Fig. 1-7) and watch for the BLUE reduced
mode indicator to illuminate (Fig. 1-8). If the window is not positioned
correctly, or is moved from the closed position, the BLUE indicator will
be extinguished. When the BLUE reduced mode indicator is not
illuminated, the unit will not operate in reduced mode, and the optional
UV light will not function.
1300 Series A2 1-7
Thermo Scientific
Section 1
Description
Use of the Window
(continued)
Figure 1-5.
250mm Opening (Work Position)
Figure 1-6.
Lighted Green LED
Figure 1-7.
Fully Closed Position (UV, if applicable)
Figure 1-8.
Lighted Blue LED
dimple
dimple
Содержание 1300 Series
Страница 25: ...Thermo Scientific Section 2 Installation 2 8 1300 Series A2 ...
Страница 31: ...Thermo Scientific Section 3 Start Up 3 6 1300 Series A2 ...
Страница 39: ...Thermo Scientific Section 4 Operation 4 8 1300 Series A2 ...
Страница 45: ...Thermo Scientific Section 5 Cleaning and Decontamination 5 6 1300 Series A2 ...
Страница 69: ...Thermo Fisher Scientific 401 Millcreek Road Marietta Ohio 45750 United States www thermofisher com ...
Страница 70: ...1300 系列 A2 II级 A2型 生物安全柜 操作手册 5991355 ...
Страница 91: ...Section 2 安装 维修阀安装 续 维修阀 维修阀插头 垫片 螺母 图2 4 阀门总成和安装 2 6 1300 Series A2 Thermo Scientific 螺纹管 底座 ...
Страница 129: ...仪器类型 部件号 序列号 维修号 地点 操作者标注 技术员 公司 记录 日期 签名 Thermo Scientific 1300 Series A2 9 1 ...