Inflow Area:
• 1.2 m models 1344/1374: Front opening 0.3 m
2
• 1.8 m models 1349/1379: Front opening 0.45 m
2
Calculation:
• Inflow Velocity (V1) = Inflow Air Volume (V2)/Actual Inflow Area (A2)
Acceptance:
• 0.5 m/s, ±0.025 m/s
Correction measures:
• Enter the service level and activate S2, change S2 until the desired
velocity is reached and save the new value.
• If the velocity can't be adjusted due to clogging of the filters, change the
exhaust filter.
• If the filter is in order, change the exhaust blower.
Note
Check of alarm limits: If S1 or S2 are changed, the alarm limits S3
and S4 must be checked and saved again.
s
Inflow Velocity – Constricted Window Method (Secondary)
Description:
• Inflow velocity measured over several points through a constricted
window.
Equipment:
• Thermal Anemometer, or equivalent (hotwire)
• Anemometer Probe Holder
8-6
1300 Series A2
Thermo Scientific
Section 8
Certification Testing
Testing Information
(continued)
Содержание 1300 Series
Страница 25: ...Thermo Scientific Section 2 Installation 2 8 1300 Series A2 ...
Страница 31: ...Thermo Scientific Section 3 Start Up 3 6 1300 Series A2 ...
Страница 39: ...Thermo Scientific Section 4 Operation 4 8 1300 Series A2 ...
Страница 45: ...Thermo Scientific Section 5 Cleaning and Decontamination 5 6 1300 Series A2 ...
Страница 69: ...Thermo Fisher Scientific 401 Millcreek Road Marietta Ohio 45750 United States www thermofisher com ...
Страница 70: ...1300 系列 A2 II级 A2型 生物安全柜 操作手册 5991355 ...
Страница 91: ...Section 2 安装 维修阀安装 续 维修阀 维修阀插头 垫片 螺母 图2 4 阀门总成和安装 2 6 1300 Series A2 Thermo Scientific 螺纹管 底座 ...
Страница 129: ...仪器类型 部件号 序列号 维修号 地点 操作者标注 技术员 公司 记录 日期 签名 Thermo Scientific 1300 Series A2 9 1 ...