Install the service valve(s) by
removing the centers of the plug
(brass) or piercing it at the
pre-determined location (plastic).
Installation spacing on side of
unit is indicated in Figure 2-3.
Note
The plugs must only be
used for the installation of service
valves in accordance with the
applicable national regulations.
s
At each side panel, service valves can be installed through the four service
valve port plugs.
1. Move the front window to the fully open position.
2. Turn the unit off and disconnect it from the power source.
3. Score the plug circularly at the pre-determined point (inside and
outside). Remove the centers.
4. Referring to Figure 2-4 on the following page, slide the gasket and the
bezel onto the threaded pipe of the service valve. Apply a thin bead of
silicon to the inner flange of the service valve.
5. From inside the sample chamber, slide the service valve through the
desired plug.
6. From the outside, slide the other gasket and the washer onto the
service valve threaded pipe and apply a thin bead of silicon to the outer
flange of the service valve.
7. Secure the service valve to the side panel using the nut.
8. Establish the connection to the supply line using a union nut.
Note
If a service valve is removed, the service valve port plug should be
re-installed.
s
1300 Series A2 2-5
Thermo Scientific
Section 2
Installation
Service Valve
Connections
Figure 2-3.
Valve Location
Service valve
port plugs
2” (5.1 cm)
5” (12.7 cm)
4”
(10.2 cm)
10.2”
(25.9 cm)
8.9”
(22.6 cm)
5.5”
(14 cm)
5”
(12.7 cm)
SmartPort
Содержание 1300 Series
Страница 25: ...Thermo Scientific Section 2 Installation 2 8 1300 Series A2 ...
Страница 31: ...Thermo Scientific Section 3 Start Up 3 6 1300 Series A2 ...
Страница 39: ...Thermo Scientific Section 4 Operation 4 8 1300 Series A2 ...
Страница 45: ...Thermo Scientific Section 5 Cleaning and Decontamination 5 6 1300 Series A2 ...
Страница 69: ...Thermo Fisher Scientific 401 Millcreek Road Marietta Ohio 45750 United States www thermofisher com ...
Страница 70: ...1300 系列 A2 II级 A2型 生物安全柜 操作手册 5991355 ...
Страница 91: ...Section 2 安装 维修阀安装 续 维修阀 维修阀插头 垫片 螺母 图2 4 阀门总成和安装 2 6 1300 Series A2 Thermo Scientific 螺纹管 底座 ...
Страница 129: ...仪器类型 部件号 序列号 维修号 地点 操作者标注 技术员 公司 记录 日期 签名 Thermo Scientific 1300 Series A2 9 1 ...