According to YY 0569-2005, tests related to the safety of the personnel,
product and environment that must be conducted on-site for Class II,
Type A2 biological safety cabinets, at a minimum frequency of annually,
are;
• downflow velocity profile test
• inflow velocity test
• airflow smoke patterns test
• HEPA filter leak test
• site installation assessment tests
– alarm functions
– exhaust system performance (proper exhaust duct negative pressure
and canopy performance).
As described in the standard, in addition to the above, the following tests
performed at the request of the customer, or at the discretion of the
certification provider.
• comfort and safety tests
– lighting intensity
– noise level
– vibration
– electrical leakage, ground circuit resistance and polarity tests
Note
Service with costs: Unless certification was expressly called for in the
specification, quotes and/or purchase order, the cost for this on-site testing
is to be paid for by the customer.
s
8-2
1300 Series A2
Thermo Scientific
Section 8
Certification Testing
Testing
Содержание 1300 Series
Страница 25: ...Thermo Scientific Section 2 Installation 2 8 1300 Series A2 ...
Страница 31: ...Thermo Scientific Section 3 Start Up 3 6 1300 Series A2 ...
Страница 39: ...Thermo Scientific Section 4 Operation 4 8 1300 Series A2 ...
Страница 45: ...Thermo Scientific Section 5 Cleaning and Decontamination 5 6 1300 Series A2 ...
Страница 69: ...Thermo Fisher Scientific 401 Millcreek Road Marietta Ohio 45750 United States www thermofisher com ...
Страница 70: ...1300 系列 A2 II级 A2型 生物安全柜 操作手册 5991355 ...
Страница 91: ...Section 2 安装 维修阀安装 续 维修阀 维修阀插头 垫片 螺母 图2 4 阀门总成和安装 2 6 1300 Series A2 Thermo Scientific 螺纹管 底座 ...
Страница 129: ...仪器类型 部件号 序列号 维修号 地点 操作者标注 技术员 公司 记录 日期 签名 Thermo Scientific 1300 Series A2 9 1 ...