Testing
Tests include factory final test, type test, field test and maintenance
test according to YY 0569-2005.
•
Factory Final Tests
The testing would be conducted before delivery in accordance with the
clauses 5.1, 5.2, 5.3, 5.4.1~5.4.5, 5.4.7~5.4.15 on YY0569-2005
•
Type Tests
According to clause 7.3 of YY0569-2005, Type test shall be randomly
sampled from cabinets for delivery. Samples for type test shall be
provided by the manufacturer, and the test must be performed by
qualified certification bodies.
The testing would be conducted in accordance with the clauses 5.1,
5.2, 5.3, 5.4 on YY0569-2005
•
Field Tests
Field testing would be performed after installation and cabinet
re-location. The testing would be conducted in accordance with the
clauses 5.1, 5.4.2, 5.4.7~5.4.9, 5.4.14 on YY 0569-2005
•
Maintenance Tests
The testing would be conducted in accordance with the clauses 5.1,
5.4.2, 5.4.7~5.4.9, 5.4.14 on YY 0569-2005 at least once a year. Test
also would be performed after filters are changed and interior parts are
repaired.
1300 Series A2
8-3
Thermo Scientific
Section 8
Certification Testing
Testing (continued)
Содержание 1300 Series
Страница 25: ...Thermo Scientific Section 2 Installation 2 8 1300 Series A2 ...
Страница 31: ...Thermo Scientific Section 3 Start Up 3 6 1300 Series A2 ...
Страница 39: ...Thermo Scientific Section 4 Operation 4 8 1300 Series A2 ...
Страница 45: ...Thermo Scientific Section 5 Cleaning and Decontamination 5 6 1300 Series A2 ...
Страница 69: ...Thermo Fisher Scientific 401 Millcreek Road Marietta Ohio 45750 United States www thermofisher com ...
Страница 70: ...1300 系列 A2 II级 A2型 生物安全柜 操作手册 5991355 ...
Страница 91: ...Section 2 安装 维修阀安装 续 维修阀 维修阀插头 垫片 螺母 图2 4 阀门总成和安装 2 6 1300 Series A2 Thermo Scientific 螺纹管 底座 ...
Страница 129: ...仪器类型 部件号 序列号 维修号 地点 操作者标注 技术员 公司 记录 日期 签名 Thermo Scientific 1300 Series A2 9 1 ...