The biological safety cabinet can be turned off (OFF mode) by pressing
and holding the ON key for five seconds until all LEDs are off.
UV mode
means the UV light is on with a set timer running. The window
must be in the closed position to protect against UV radiation. The light
will not turn on with the window open.
The yellow LED “UV disinfection routine activated” is lighted until the
preset time has elapsed. Then, the UV lights turn off automatically, and
the status indicator turns off.
The sample chamber illumination is disabled in this mode, as well as the
internal power source in the sample chamber.
1. Move the front window to the maximum opening position. This
automatically switches the blowers to full speed.
2. Install needed work materials within the sample chamber work area.
Avoid blocking the front air intake grille.
3. Move the front window to the work position (the green status
indicator “Front Window in Work” position is illuminated) and wait
until the airflow has stabilized (the green status indicator “Airflow
Steady” is illuminated).
Warning
Personal and product protection is ensured only if the airflow
system of the device is working properly. If alarm system issues a failure
message for more than a few minutes while the front window is in the
work position, stop all applications that may jeopardize worker safety.
s
4. Load the work tray with samples.
5. For work breaks or for extended experimental phases without manual
intervention, switch the device to standby mode by closing the
window.
4-4
1300 Series A2
Thermo Scientific
Section 4
Operation
Operating Modes
(continued)
Loading the Chamber
Содержание 1300 Series
Страница 25: ...Thermo Scientific Section 2 Installation 2 8 1300 Series A2 ...
Страница 31: ...Thermo Scientific Section 3 Start Up 3 6 1300 Series A2 ...
Страница 39: ...Thermo Scientific Section 4 Operation 4 8 1300 Series A2 ...
Страница 45: ...Thermo Scientific Section 5 Cleaning and Decontamination 5 6 1300 Series A2 ...
Страница 69: ...Thermo Fisher Scientific 401 Millcreek Road Marietta Ohio 45750 United States www thermofisher com ...
Страница 70: ...1300 系列 A2 II级 A2型 生物安全柜 操作手册 5991355 ...
Страница 91: ...Section 2 安装 维修阀安装 续 维修阀 维修阀插头 垫片 螺母 图2 4 阀门总成和安装 2 6 1300 Series A2 Thermo Scientific 螺纹管 底座 ...
Страница 129: ...仪器类型 部件号 序列号 维修号 地点 操作者标注 技术员 公司 记录 日期 签名 Thermo Scientific 1300 Series A2 9 1 ...