86
Example
SWEEP:GRAP:SHOW ON
Displays the sweep graph on the LCD display.
4.9.4 SWEEP :GRAPh:LINE
Set
Query
Description
Sets or returns which lines are shown on the sweep graph.
Syntax
Query Syntax
SWEEP:GRAPh:LINE <NR1>
SWEEP:GRAPh:LINE?
Parameter/ Return
parameter
<NR1> Description
0
Turn all lines off/all lines are off.
1
Displays the graph line for the primary test item. See
page 41 for details.
For example: V for ACW, DCW and GB tests, I for IR
tests.
2
Displays the graph line for the secondary test items.
For example: I for ACW and DCW tests, R for IR and
GB tests.
3
Turn all lines on/all lines are on.
Example
SWEEP:GRAP:LINE 3
Turns all the graph lines on.
4.9.5 SWEEP:STARt:TIME
Set
Query
Description
Sets or returns the start time (STA.t) of the sweep graph in
milliseconds.
This setting will also set what the time for first point will be for
the sweep data that is returned in the SWEEP:DATA:SHOW
query.
Syntax
Query Syntax
SWEEP:STARt:TIME <NR2>
MANU:RTIMe?
Parameter/ Return
parameter
<NR2>
0.1~1999.8 milliseconds
Example
SWEEP:STARt:TIME 100.0
Sets the sweep start time to 1 second.
4.10 Auto Commands
4.10.1 AUTO:STEP
Set
Query
Description
Sets or queries the AUTO number (automatic test number).
Syntax
Query Syntax
AUTO:STEP <NR1>
AUTO:STEP?
Parameter/ Return
parameter
<NR1>
1~100.
Example
AUTO:STEP 100