MANU Utility
Settings
The Manu Utility settings are configured for each MANU
test separately. The settings include: ARC MODE, PASS
HOLD, FAIL MODE, MAX HOLD and GROUND MODE.
L I T Y
I
M A U
N
U T
2
M A N U = * * * - 0 0
A R C M D E : O
O
F
F
G R O U N D M O D E :
N
O
P A S S H
L D :
O
O F
F
F A I L M D E :
O
P
S T O
M A X H O L D : O F F
2.2 Test Lead Connection
This section describes how to connect the STW-9000 to a DUT for withstanding,
insulation resistance or ground bond testing.
2.2.1 ACW, DCW, IR Connection
Background
ACW, DCW and IR tests use the HIGH VOLTAGE terminal
and RETURN terminal with the GHT-114 test leads.
ACW, DCW, IR
Connection
High Voltage
terminal
DUT
Return terminal
Steps
1.
Turn the power off on the safety tester.
2.
Connect the high voltage test lead(red) to the HIGH
VOLTAGE terminal and screw firmly into place.
3.
Connect the return test lead(white) into the RETURN
terminal and screw the protector bar into place, as shown
below.
HIGH VOLTAGE
Terminal
RETURN
Terminal