6.2 Error Messages
6.2.1 System Self-Test
The following error messages or messages may appear on the STW screen during
the Start-Up initialization. If any of these error messages appear on the STW-9000,
please see an authorized TEXIO distributor.
Error Messages
Description
0x11
EEPROM1 Error
0x12
EEPROM1 Error
0x21
W-V Offset Error (W-V: ACW/DCW voltage)
0x22
W-I Offset Error (W-I: ACW/DCW current)
0x23
IR-I Offset Error
0x24
GB-I Offset Error
6.2.2 Test Errors
The following error messages or messages may appear on the STW screen when
configuring or running tests.
Error Messages
Description
TIME ERR
TIME ERR is displayed For ACW tests.
STW-9900:
HI SET ≥ 80.00mA~100.0mA
RAMP
time + TEST TIME setting is > 240 sec
STW-9800:
HI SET ≥ 30.00mA~40.0mA
RAMP
time + TEST TIME setting is > 240 sec
OVER 100W
(STW-9900)
For DCW tests. OVER 100W is displayed if the HI
SET setting multiplied by the Voltage setting is greater
than 100W
OVER 50W
(STW-9800)
For DCW tests. OVER 50W is displayed if the HI SET
setting multiplied by the Voltage setting is greater than
50W
I ERR
For ACW, DCW tests. Shown when the current is set
too high.
SHORT
Voltage is too low or there is no High Voltage output.
Indicates that the DUT could be shorted.
V ERR
For ACW, DCW tests. Indicates that an abnormal
voltage has been detected.
V = 0
For GB tests. Voltage is equal to 0. Check to see that
the SENSE H or SOURCE H is not open.
R ERR
For IR tests. The voltage is too high or resistance=0Ω.
Check to see whether the DUT or test lead is shorting.
For GB tests. The resistance is too high.