4.9.2 SWEEP<X>:DATA:SHOW
Query
Description
Returns the data associated with a sweep graph.
Data can be returned in one of two ways; either all the data
can be returned or only the data at a particular point in time.
The test points are evenly distributed. There can be up to 190
data points.
If only the data from a single point is returned then the data is
returned in the following format*:
DATA POINT, VSET, ISET, TIME, CR+LF
If the all the data for the all the points is returned then the data
is returned in the following format*:
ACW MODE,CR+LF
No.,V(kV),I(mA), T(S) ,CR+LF
001,0.071,0.032,0000.1,CR+LF
002,0.111,0.047,0000.2,CR+LF
…………………………………
013,0.601,0.215,0001.3,CR+LF
END
*Where CR+LF is a carriage return and line feed code. Time
is in seconds.
Query Syntax
SWEEP<X>:DATA:SHOW?
Parameter
<X>
1~190 (single data point)
0 (all data points)
Single Data Point
Example
SWEEP10:DATA:SHOW?
> 010,0.106,00.00,0001.0, CR+LF
Returns the data at point 10, which is at the 1 second time for
the sweep test.
All Data Points
Example
SWEEP0:DATA:SHOW?
>ACW MODE,CR+LF
>No.,V(kV),I(mA), T(S) ,CR+LF
>001,0.071,0.032,0000.1,CR+LF
>002,0.111,0.047,0000.2,CR+LF
>…………………………………
>013,0.601,0.215,0001.3,CR+LF
>END
This will return all the data from the sweep graph.
4.9.3 SWEEP:GRAPh:SHOW
Set
Query
Description
Turns the sweep graph on or off on the STW-9000 display.
Syntax
Query Syntax
SWEEP:GRAPh:SHOW {ON|OFF}
SWEEP:GRAPh:SHOW?
Parameter/ Return
parameter
ON
Turn the sweep graph on.
OFF
Turn the sweep graph off.