26
3.
Use the scroll wheel to set the ARC
mode.
ARC MODES: OFF, ON AND CONTINUE,
ON AND STOP
4.
Press the EDIT/SAVE key to save and
exit the MANU Utility and go back to
EDIT status.
EDIT/SAVE
Note
The ESC key can be pressed at any time in the Utility
menu to cancel and exit.
5.
If the ARC MODE was set to either ON AND CONTINUE,
or ON AND STOP, the ARC current level can be edited.
6.
Use the UP/DOWN arrow keys to move
the cursor to the ARC setting.
7.
Use the scroll wheel to edit the ARC
level.
ACW
DCW
2.000mA~200.0mA
2.000mA~040.0mA
Note
The ARC setting range is directly related to the HI SET
current limit.
ACW:
HI SET Limit
ARC Range
0.001mA~1.100mA
2.000mA
01.11mA~11.00mA
02.00mA ~20.00mA
011.1mA~110.0mA
002.0mA ~200.0mA
DCW:
HI SET Limit
ARC Range
0.001mA~1.100mA
2.000mA
01.11mA~11.00mA
02.00mA ~20.00mA
011.1mA~021.0mA
002.0mA ~040.0mA
2.3.12 Setting PASS HOLD
Background
The PASS HOLD settings only apply to the selected test in
an AUTO test. When the PASS HOLD setting is set to ON,
a PASS judgment is held until the START button is
pressed.
Note
The PASS HOLD setting only applies to AUTO tests. This
setting is ignored when running
a single
MANU test.