73
Parameter
(MANU mode)
No parameter needed for MANU mode.
Parameter
(AUTO mode)
<x>
<NR1>1~16. Step number.
Return parameter <string>
Returns the test status of the test in the
following format:
function, judgment or status, test voltage,
test current or resistance, test time or ramp
time
Function
ACW, DCW, IR, GB
Judgment
/Status
PASS, FAIL
VIEW
Test voltage
unit
Test current
/Test resistance
unit
resunit
Test time
/Ramp time
T=time+S
R=time+S
Example
(in MANU mode)
MEAS?
>ACW, FAIL , 0.024kV ,0.013 mA ,R=000.1S
Returns the test result of the current manual test.
Example
(in AUTO mode)
MEAS10?
>IR, FAIL ,0.225kV ,999M ohm,T=010.3S
Returns step 10 of the current automatic result.
4.7.3 MAIN:FUNCtion
Set
Query
Description
Changes the mode between AUTO and MANU.
Syntax
Query Syntax
MAIN:FUNCtion {MANU|AUTO}
MAIN:FUNCtion ?
Parameter/ Return
parameter
MANU
Puts the tester mode to MANU.
AUTO
Puts the tester mode to AUTO.
Example
MAIN:FUNC MANU
Sets the tester to MANU mode.
4.8 Manual Commands
4.8.1 MANU:STEP
Set
Query
Description
Sets the MANU test number.
Syntax
Query Syntax
MANU:STEP <NR1>
MANU:STEP?
Parameter/ Return
parameter
<NR1>
0~100.
Example
MANU:STEP 100
Sets the manual test number to 100.