72
4.6.8 SYSTem:GPIB:VERSion
Query
Description
Queries the GP-IB version.
Query Syntax
SYSTem:GPIB:VERSion?
Return parameter <string>
Returns:
The GP-
IB version as a string “GPIB,V1.00”
or
“No GPIB connected” if there is not a GP-IB
device configured/connected.
Query Example
SYST:GPIB:VERS?
>GPIB,V1.00
Returns the GP-IB version.
4.7 Function Commands
4.7.1 FUNCtion:TEST
Set
Query
Description
Turns the currently selected test (output) on or off.
When HOLD is displayed on the screen during AUTO tests, use
the FUNCtion:TEST command to move on to the next step.
Setting the FUNCtion:TEST command to OFF at the end of a
test will also temporarily turn the PASS/FAIL buzzer sound off.
Syntax
Query Syntax
FUNCtion:TEST {ON|OFF}
FUNCtion:TEST?
Parameter
ON
Turns the test on.
OFF
Turns the test off.
Return parameter TEST ON
Test is on.
TEST OFF
Test is off.
Example
FUNC:TEST ON
Turns the output on.
4.7.2 MEASure<x>
Query
Description
Returns the test parameters & results of the tester in either
MANU or AUTO mode.
MANU mode: Returns the test parameters & results of a MANU
test.
AUTO mode: Returns the test parameters & results of the
selected step (1-16) of the AUTO test.
Return parameters: function, judgment/status, test voltage, test
current/resistance, test time (time of completed test) or ramp
time (elapsed time of test that has not been completed.
Query Syntax
MEASure<x>?