77
Example
MANU:ACW:REF 0.01
Sets the ACW reference to 0.01 mA.
4.8.11 MANU:ACW:ARCCurrent
Set
Query
Description
Sets or returns the ACW ARC current value in mA. ARC must
be enabled before the ARC current can be set. The test must
first be in ACW mode before this command can be used.
ARC current uses the same range as the HI SET value. The
ARC current is limited to 2X the HI SET value.
Syntax
Query Syntax
MANU:ACW:ARCCurrent <NR2>
MANU:ACW:ARCCurrent?
Parameter/ Return
parameter
<NR2>
2.000 ~ 200.0
Example
MANU:ACW:ARCC 0.04
Sets the ACW ARC value to 0.04 mA.
4.8.12 MANU:DCW:VOLTage
Set
Query
Description
Sets or returns the DCW voltage in kV. The test must first be
in DCW mode before this command can be used.
Note: A “DC Over 100W” error will result if the DCW Voltage X
HI SET value is > 100 watts.
Syntax
Query Syntax
MANU:DCW:VOLTage <NR2>
MANU:DCW:VOLTage?
Parameter/ Return
parameter
<NR2>
0.100 ~ 6.100 (kV)
Example
MANU:DCW:VOLT 6
Sets the DCW voltage to 6 kV.
4.8.13 MANU:DCW:CHISet
Set
Query
Description
Sets or returns the DCW HI SET current value in milliamps.
The test must first be in DCW mode before this command can
be used.
Note: A “DC Over 100W” error will result if the DCW Voltage X
HI SET value is > 100 watts.
Syntax
Query Syntax
MANU:DCW:CHISet <NR2>
MANU:DCW:CHISet?
Parameter/ Return
parameter
<NR2>
0.001 ~ 21.0
Example
MANU:DCW:CHIS 5
Sets the DCW CHI SET current to 5mA.