Table 1: Output Limitation in Withstanding Voltage Testing
STW-9800 Upper Current
Pause
Output Time
AC
30mA
≤ I ≤ 40mA At least as long as
the output time
Maximum 240 seconds
0.001mA
≤ I < 30mA Not necessary
Continuous output possible
DC
0.001mA
≤ I ≤ 10mA Not necessary
Continuous output possible
STW-9900 Upper Current
Pause
Output Time
AC
80mA
≤ I ≤ 100mA At least as long as
the output time
Maximum 240 seconds
0.001mA
≤ I < 80mA Not necessary
Continuous output possible
DC
0.001mA
≤ I ≤ 20mA Not necessary
Continuous output possible
GB
15A < I
≤ 32A
Maximum 999.9 seconds
3A
≤ I ≤ 15A
Not necessary
Maximum 999.9 seconds
NOTE: Output Time = Ramp Time + Test Time.