4.8.14 MANU:DCW:CLOSet
Set
Query
Description
Sets or returns the DCW LO SET current value in milliamps.
The LO SET value must be less than the HI SET value. The
test must first be in DCW mode before this command can be
used.
The LO SET range must use the HI SET range. If all the digits
in the LO SET range are outside the HI SET range, an error
will be produced. All digits outside the HI SET range are
ignored and will not be used.
For example:
HI SET value: 12.34
LO SET value1: 0.005
error
LO SET value2: 0.053
no error
In the example above LO SET value1 will produce an error as
all digits are outside the range of HI SET. LO SET value2 will
not produce an error, but will return 0.05, not 0.053.
Syntax
Query Syntax
MANU:DCW:CLOSet<NR2>
MANU:DCW:CLOSet?
Parameter/ Return
parameter
<NR2>
0.000 ~ 20.9
Example
MANU:DCW:CLOS 2.00
Sets the DCW LO SET current to 2mA.
4.8.15 MANU:DCW:TTIMe
Set
Query
Description
Sets or returns the DCW test time in seconds. The test must
first be in DCW mode before this command can be used.
In special MANU mode, the TIMER can be turned off.
Syntax
Query Syntax
MANU:DCW:TTIMe {<NR2>|OFF}
MANU:DCW:TTIMe?
Parameter
<NR2>
OFF
0.5 ~ 999.9 seconds
TIMER OFF (special MANU mode).
Return parameter <NR2>
TIME OFF
0.5 ~ 999.9 seconds
TIMER is OFF (special MANU mode).
Example
MANU:DCW:TTIM 1
Sets the DCW test time to 1 second.