31
2-1-4. CV Mode
Description
In Constant Voltage Mode, the unit will maintain
a constant voltage. In CV mode you set the
constant voltage level. For more details on CV
mode, see the appendix on page 136.
Warning
If you change the mode or the range when the
load is already on, the load will be turned off
automatically.
Operation
1.
Make sure the load is off.
2.
Press
Main
.
3.
Select CV mode with the
Mode[F1]
soft-key.
4.
Select the current range with the
I Range[F2]
soft-key.
Range:
High, Middle, Low
5.
Select the voltage range with the
V Range[F3]
soft-key.
Range:
High, Low
6.
Set the voltage level parameters using the scroll
wheel and number pad.
Set CV A Value and/or CV B Value.
The maximum and minimum voltage levels
depend on the selected voltage range.
7.
Set the remaining basic configuration settings
such as the response settings. See page 37 for
details.
Display
Mode
CV
I Range
H 35A
V Range
L 15V
Response
Slow
Configure
0.000
V
0.000
A
0.00
w
CV B Value
15.000
CV A Value
15.000
LOAD
RS232
A Value
Fine
31/Oct/2013
V
V
Mode
Current range
Voltage
range
Active
setting
Current
setting
Voltage
settings
Note
Basic CV mode configuration is complete. See
page 37 for more configuration options.
The current range and voltage range applies to
all the operating modes.