143
Resolution
H Range
10mV
L Range
1mV
Accuracy of Setting
*1
H, L Range ±(0.1 % of set + 0.1 % of f.s.)
Input current variation
*2
H Range
50mV
L Range
12mV
*1 At the sensing point during remote sensing under the operating
range of the input voltage. It is also applied for the condition of
the parallel operation.
*2 With respect to a change in the current of 10 % to 100 % of the
rating at an input voltage of 1.5 V (during remote sensing).
7-7-6. CP Mode
Model
LSG-175
LSG-350
LSG-1050
Operating Range
H Range
17.5W~175W
35W~350W
105W~1050W
M Range
1.75W~17.5W
3.5W~35W
10.5W~105W
L Range
0.175W~1.75W 0.35W~3.5W
1.05W~10.5W
Setting Range
H Range
0W~178.5W
0W~357W
0W~1071W
M Range
0W~17.85W
0W~35.7W
0W~107.1W
L Range
0W~1.785W
0W~3.57W
0W~10.71W
Resolution
H Range
10mW
10mW
100mW
M Range
1mW
1mW
10mW
L Range
0.1mW
0.1mW
1mW
Accuracy of Setting
*1
±(0.6 % of set + 1.4 % of f.s.
*2
)
+ Vin
*3
/ 500kΩ
*1 It is not applied for the condition of the parallel operation.
*2 M range applies to the full scale of H range.
*3 Vin = Input terminal voltage of electric load.
7-7-7. Slew Rate
Model
LSG-175
LSG-350
LSG-1050
Setting Range (CC Mode)
H Range
2.5mA/us~2.5A/us 5mA/us~5A/us
16mA/us~16A/us
M Range
250uA/us
~250mA/us
500uA/us~500mA/us 1.6mA/us~1.6A/us
L Range
25uA/us~25mA/us 50uA/us~50mA/us
160uA/us~160mA/us
Содержание LSG SERIES
Страница 8: ......
Страница 23: ...11 GRA 414 E EIA standard 132 382 5 477 6 GRA 414 J JIS standard 149 382 5 477 6 ...
Страница 149: ...137 7 6 Operating Area LSG 175 ...
Страница 150: ...138 LSG 350 ...
Страница 151: ...139 LSG 1050 ...
Страница 166: ......