Application Information
Picture Application
189
PRISM MPI-MPX User Manual
To Adjust the Lightmeters Cursors
•
To move the Lightmeter cursors, select any part of the crosshairs or the
measurement numbers and drag the target to the new location. The measurement
updates automatically.
•
To remove a selected Lightmeter cursor from the Picture, unselect the color in the
Settings menu.
•
To remove all the Lightmeters cursors from the Picture, unselect all of the
Lightmeter colors or set the Lightmeter Overlay to Off.
Luminance False Color
The Luminance False Color display replaces pixels that fall within a specified range of
nits or stops values to a color that you specify. Up to 9 configurable ranges can be
defined. For example, you can use the Luminance False Color display to tag pixels that
are within a specified HDR range.
(Copyright 2008, Blender Foundation / www.bigbuckbunny.org)
The false color ranges are defined in nits or stops depending on the gamma curve set
up in the input (select Settings and then Input). If a camera-based gamma curve is set,
such as S-Log2, the false color ranges are in stops. If a display-based gamma curve is
set, such as SDR Narrow, the false color ranges are in Nits.
These controls adjust the Luminance False Color display:
Meter Overlay
Set to On to enable the Meter Overlay in the Picture application. The overlay shows the
false color definitions in nits or stops made in the Luminance False Color table at the
bottom of the menu. The overlay can be moved to the right, bottom, left, or top of the
image by dragging the meter with a mouse or by touch.
Default: By default, the false colors are defined according to the BT.2408 specifications
if a display-based gamma curve is set for the input, or a typical camera specification if a
camera-based gamma curve is set for the input. If changes are made to the false color
Содержание PRISM MPI
Страница 1: ...PRISM MPI and MPX SDI IP Waveform Monitor User Manual April 2021 D00010020C...
Страница 6: ...6 PRISM MPI MPX User Manual...
Страница 10: ...10 PRISM MPI2 25 MPX2 25 User Manual...
Страница 54: ...Getting Started Activity Center PRISM MPI MPX User Manual 54...
Страница 76: ...Methods of Operation Remote Control through API Commands PRISM MPI MPX User Manual 76...
Страница 138: ...Configure the Instrument Troubleshooting PRISM MPI MPX User Manual 138...
Страница 304: ...PTP Operational Overview Compensating for Causes of Asymmetric Delay PRISM MPI MPX User Manual 304...