Trace Nitrogen in Argon Analyzer
Operation
Teledyne Analytical Instruments
49
If instrument is turned off, the next time the instrument is powered,
the instrument will automatically perform a calibration cycle after 3
minutes of entering the sample mode if
AUTO CAL
functions were on
prior to shut down.
4.5.1.7
F
LOW
This function allows the user to set the flowrate through the flow
control valve. The first line of the display indicates the current flow
setting in ccm and can be manipulated using the
▲
/
▼
arrow keys. The
second line of the display is a read-only item that displays the live
sensor voltage within the flow control module.
To set or change the flow rate, press the
SYSTEM
key and use the
◄►
arrow keys to position the blinking over the
More
function, and
press ENTER. A second screen of system function will appear. Use the
◄►
arrow keys to position the blinking over the
Flow
function, and
press ENTER to enter the function. The following screen will appear:
Flow: 100.0 ccm
Sens Vout: 300.0 mV
The current flow rate (100) will be blinking and can be changed
using the
▲
/
▼
arrow keys. Press ENTER when the proper flow rate is
indicated on the display screen.
4.5.1.8
T
RACKING
D
URING
C
ALIBRATION
The user has the option of setting the preference as to whether the
analog outputs track the display readings during calibration or not. The
function can be toggled between Track or Hold in this function.
To set the preference, press the
SYSTEM
key and the first System
menu will appear in the VFD display. Use the
◄►
arrow keys to
position the blinking over the
More
function, and press ENTER. A
second screen of system function will appear. Use the
◄►
arrow keys to
position the blinking over the
Track (
or
Hold)
function, and toggle
between Track or Hold using the
▲
/
▼
arrow keys. Press ENTER when
the desired selection is indicated on the display screen.
If Hold is selected, the next function
Calholdtimer
sets the hold
period.
Содержание 2120XL
Страница 12: ...Model 2120XL Teledyne Analytical Instruments xii ...
Страница 28: ...Operational Theory Model 2120XL Teledyne Analytical Instruments 16 ...
Страница 46: ...Installation Model 2120XL Teledyne Analytical Instruments 34 ...
Страница 76: ...Operation Model 2120XL Teledyne Analytical Instruments 64 ...
Страница 82: ...Maintenance Model 2120XL Teledyne Analytical Instruments 70 ...