Trace Nitrogen in Argon Analyzer
Installation
Teledyne Analytical Instruments
31
3.6.2 Gas Connections to the Instrument
CAUTION:
EQUIPMENT DAMAGE MAY RESULT IF THE
ANALYTICAL CELL IN THIS UNIT IS EXPOSED TO
PRESSURE ABOVE ATMOSPHERIC, CAUSING IT TO
BREAK OR SHATTER. TO PREVENT THIS, NEVER
EXCEED 20 PSIG (138 KPA) ON THE INLET PORT.
ALWAYS ENSURE THAT THE VENT IS AT
ATMOSPHERIC PRESSURE.
DO NOT REMOVE THE FLOW CONTROL ORIFICE
FROM THE SAMPLE INLET OF THE ANALYZER.
DO NOT BLOCK THE VENT.
To connect the gas line:
1.
Connect an unobstructed vent line to the 1/8-in Swagelok
fitting on the vent port on the rear of the analyzer.
2.
Connect the inlet line from the sample manifold system,
regulated to 4 to 20 psig (41 kPa), to the 1/8-in Input port
fitting on the rear of the analyzer.
3.7 Purging
The Model 2120XL is equipped with a flow control module
installed in the analyzer sample inlet to regulate flow and pressure to the
analytical cell. A maximum inlet pressure of 20 psig (138 kPa) is
allowable.
CAUTION:
OZONE EMISSIONS ARE POSSIBLE IF AIR OR
MOISTURE IS IONIZED IN THE SAMPLING SYSTEM.
THE SAMPLE PATH MUST BE RELATIVELY FREE
OF AIR AND/OR MOISTURE BEFORE APPLYING THE
IONIZING VOLTAGE.
KEEP THE MODEL 2120XL POWER OFF AND PURGE
THE EXTERNAL TUBING. THEN APPLY POWER TO
PURGE THE INTERNAL LINES. ANALYZER POWER
IS NEEDED TO PURGE ANALYZER SINCE THE
FLOW CONTROL MODULE NEEDS POWER TO
OPERATE.
Содержание 2120XL
Страница 12: ...Model 2120XL Teledyne Analytical Instruments xii ...
Страница 28: ...Operational Theory Model 2120XL Teledyne Analytical Instruments 16 ...
Страница 46: ...Installation Model 2120XL Teledyne Analytical Instruments 34 ...
Страница 76: ...Operation Model 2120XL Teledyne Analytical Instruments 64 ...
Страница 82: ...Maintenance Model 2120XL Teledyne Analytical Instruments 70 ...