Operation Model
2120XL
Teledyne Analytical Instruments
48
Model 2120XL is accurate to 2-3 %. Accordingly, internally
scheduled calibrations can vary 2-3 % per day.
To setup an Auto Cal cycle:
Choose
SYSTEM
from the function buttons. Use the
◄►
arrow
keys to position the blinking over the
More
function, and press ENTER.
A second display of
System
functions will appear. Use the
◄►
arrow
keys again to position the blinking over the
Auto Cal
function, and
press ENTER to enter the function. The following screen will appear:
Zero In 0d 0h Off
Span In 0d 0h Off
Press the
▲
/
▼
arrow keys to blink
ZERO
(or
SPAN
), then press
ENTER
again.
Note: You will be unable to set
OFF
to
ON
if a zero interval is
entered.
A
Span Every ...
(or
Zero Every ...
) screen appears.
Zero schedule: OFF
Day: Ød Hour: Øh
Move the blinking highlight to the
Day
field , then use the
▲
/
▼
arrow keys to display the correct value for the day.
Once the correct day is displayed press ENTER to accept the value
and to move to the start-time value. Use the
▲
/
▼
arrow keys to set a
start-time value. Press ENTER again to accept the value.
To turn ON the
SPAN
and/or
ZERO
cycles (to activate the auto cal
feature):
If a scheduled day and time has just been entered, press ESCAPE to
return to the previous menu or press
SYSTEM
again and move to the
AUTO CAL
function with the
◄►
keys
and press ENTER
again.
When the
ZERO/SPAN
values screen appears, use the
▲
/
▼
arrows
to blink the
ZERO
(or
SPAN
) and press ENTER
to go to the next screen.
Use the
◄►
arrows to select the
OFF/ON
field. Use the
▲
/
▼
arrows to
set the
OFF/ON
field to
ON
. You can now turn these fields ON because
there is a nonzero span interval defined.
Содержание 2120XL
Страница 12: ...Model 2120XL Teledyne Analytical Instruments xii ...
Страница 28: ...Operational Theory Model 2120XL Teledyne Analytical Instruments 16 ...
Страница 46: ...Installation Model 2120XL Teledyne Analytical Instruments 34 ...
Страница 76: ...Operation Model 2120XL Teledyne Analytical Instruments 64 ...
Страница 82: ...Maintenance Model 2120XL Teledyne Analytical Instruments 70 ...