
Trace Nitrogen in Argon Analyzer
Operation
Teledyne Analytical Instruments
41
Weight of Digital
Filter: 4
The current filter setting (4 in this case) will be flashing. It can be
modified from 0 (no filtering) to 10 maximum filtering using the
▲
/
▼
keys. Pressing ENTER will accept the value and return to the
SYSTEM
menu.
The default setting should suffice for most applications. In some
applications where speeding the response time with some trade off in
noise is of value, the operator could decrease the number of the digital
filter. In applications where the signal is noisy, switch to a higher
number, however, the response time will decrease.
4.5.1.2
S
ELF
-T
EST
The system self-test is automatically initiated when the unit is
powered up. It can also be activated by the user from the
SYSTEM
menu.
To manually initiate a self-test, press the STSTEM key and navigate to
Self-Test using the
◄►
arrow keys. Press ENTER to engage the
function and the following screen appears:
Perform System Test?
<ENT> Yes <ESC> No
Pressing ENTER will begin the Self-Test. A successions of
diagnostic screens will follow:
RUNNING DIAGNOSTIC
Testing Preamp — 83
During preamp testing there is a countdown in the lower right
corner of the screen. When the testing is complete, the results are
displayed.
Power: ##
Analog: ##
Preamp: ##
The module is functioning properly if it is followed by
OK
. If a
problem is detected, a number will appear next to the heading and
Содержание 2120XL
Страница 12: ...Model 2120XL Teledyne Analytical Instruments xii ...
Страница 28: ...Operational Theory Model 2120XL Teledyne Analytical Instruments 16 ...
Страница 46: ...Installation Model 2120XL Teledyne Analytical Instruments 34 ...
Страница 76: ...Operation Model 2120XL Teledyne Analytical Instruments 64 ...
Страница 82: ...Maintenance Model 2120XL Teledyne Analytical Instruments 70 ...