SP3X Sensors
LXT-380 Universal Transmitter
Teledyne Analytical Instruments
64
Fluoride Electrodes
a.
Moisten the end of the Q-tip with water and dip it in the alumina
polishing powder to pick up a small amount of the powder.
b.
Rub the polishing powder onto the fluoride crystal in a circular
motion and moisten the tip if necessary to produce a liquid
consistency more than a paste.
c.
Polish the electrode for about 30 seconds and examine the tip for
a shiny surface, repeat if necessary.
d.
Rinse the electrode with distilled water and soak the electrode in
a low ppm Calibration solution for a few minutes.
e.
Perform a Two Point Calibration.
Other Solid State electrodes, Bromide, Cadmium, Chloride,
Copper, Lead, Sulfide, Silver…
a.
Place a few drops of distilled water on the blue polishing strip to
wet the polishing surface.
b.
Hold the electrode with the sensing surface facing up.
c.
Slide the polishing strip back and forth across the electrode tip,
the sensing tip will be abraded and a new electrode surface will
be generated.
d.
Polish the electrode for about 30 seconds and examine the tip for
a shiny surface, repeat if necessary.
e.
Rinse the electrode with distilled water and soak the electrode in
a low ppm Calibration solution for a few minutes.
f.
Perform a Two Point Calibration.
8.3.7 D
ISSOLVED
O
XYGEN
C
ARTRIDGES
The Teflon membrane of the Dissolved Oxygen sensors is fragile
and should be cleaned using a soft cloth or an artist style paint brush
while rinsing with a stream of water. Dilute dish washing detergents can
be used to remove oily films. Solvents are not recommended as they
could diffuse through the membrane and harm the electrode. Strong
acid/alkaline solutions should not harm the electrode but should only be
used as a last resort before replacing the electrode.
Содержание LXT-380
Страница 2: ......
Страница 4: ......
Страница 5: ......
Страница 6: ......
Страница 19: ...LXT 380 Universal Transmitter Teledyne Analytical Instruments xiv Blank Page...
Страница 27: ...Introduction LXT 380 Universal Transmitter Teledyne Analytical Instruments 8 1 8 APPROVALS 1 8 1 ENCLOSURE...
Страница 39: ...Operation LXT 380 Universal Transmitter Teledyne Analytical Instruments 20 Figure 3 1 LXT 380 Screen Map...
Страница 55: ...Operation LXT 380 Universal Transmitter Teledyne Analytical Instruments 36 Blank Page...
Страница 69: ...Maintenance LXT 380 Universal Transmitter Teledyne Analytical Instruments 50 Blank Page...
Страница 70: ...LXT 380 Universal Transmitter Troubleshooting Teledyne Analytical Instruments 51 6 0 TROUBLESHOOTING...
Страница 71: ...Troubleshooting LXT 380 Universal Transmitter Teledyne Analytical Instruments 52 Blank Page...
Страница 73: ...Parts and Accessories LXT 380 Universal Transmitter Teledyne Analytical Instruments 54 Blank Page...
Страница 78: ...LXT 380 Universal Transmitter SP3X Sensors Teledyne Analytical Instruments 59...
Страница 90: ...LXT 380 Universal Transmitter Appendix Teledyne Analytical Instruments 71 Appendix A Auto Cal Buffer Tables...
Страница 91: ...Appendix LXT 380 Universal Transmitter Teledyne Analytical Instruments 72 B LXT 380 Hart Menu...
Страница 92: ...LXT 380 Universal Transmitter Appendix Teledyne Analytical Instruments 73...
Страница 96: ...LXT 380 Universal Transmitter Appendix Teledyne Analytical Instruments 77...
Страница 97: ...Appendix LXT 380 Universal Transmitter Teledyne Analytical Instruments 78...
Страница 98: ...LXT 380 Universal Transmitter Appendix Teledyne Analytical Instruments 79 Fault Status WARNING STATUS...
Страница 99: ...Appendix LXT 380 Universal Transmitter Teledyne Analytical Instruments 80 Sensor Type...
Страница 101: ...Appendix LXT 380 Universal Transmitter Teledyne Analytical Instruments 82 E Drawings...
Страница 102: ...LXT 380 Universal Transmitter Appendix Teledyne Analytical Instruments 83...
Страница 103: ...Appendix LXT 380 Universal Transmitter Teledyne Analytical Instruments 84...