SP3X Sensors
LXT-380 Universal Transmitter
Teledyne Analytical Instruments
62
8.3.3 ELECTRODE CLEANING
An important aspect of sensor maintenance is the service of the
electrode cartridge. After being in operation, an electrode may begin to
exhibit slow response or non-reproducible measurements. This may be
due to coating of the measurement electrode or clogging of the reference
junction. Regular electrode cleaning reduces problems associated with
the coating and clogging. Frequency of cleaning will depend on the
process and application. The following procedures are used to clean pH
and ORP electrodes.
If possible, the electrode should be cleaned without removing it
from the sensor body. However, if the electrode must be removed, the o-
rings must be inspected and re-lubricated. See section 3.10.
8.3.4 PH ELECTRODE CARTRIDGE CLEANING
Remove the sensor from the process and carefully wash the wetted
end of the electrode cartridge in a mild solution of detergent and water
or with methyl alcohol. If the electrode response is not improved, soak
the electrode in 0.1 molar HCl for 5 minutes. Remove and rinse the
electrode with tap water and soak in 0.1 molar NaOH for 5 minutes.
Remove the electrode from the NaOH solution, rinse the electrode
and soak in a 4 pH buffer solution for 10 minutes. This should improve
the response of the electrode. If not, replace the electrode.
If the electrode must be left out of the process for an extended
period of time, store it in a solution of water saturated with KCl or a 4.0
pH buffer solution.
Teledyne does not recommend the storage of
electrodes in distilled or deionized water
.
8.3.5 ORP ELECTRODE CARTRIDGE CLEANING
Cleaning the platinum surface to remove coating can be done using
an abrasive cleaner like 600-800 grit wet/dry sand paper or chemical
reagents specific for the type of coating. Abrasive cleaning is the most
common method of cleaning and is usually sufficient to restore the
platinum surface; however, some processes can form a hard coating
requiring chemical cleaning with a strong acid solution. Acid solutions
greater than 10% are not recommended.
Содержание LXT-380
Страница 2: ......
Страница 4: ......
Страница 5: ......
Страница 6: ......
Страница 19: ...LXT 380 Universal Transmitter Teledyne Analytical Instruments xiv Blank Page...
Страница 27: ...Introduction LXT 380 Universal Transmitter Teledyne Analytical Instruments 8 1 8 APPROVALS 1 8 1 ENCLOSURE...
Страница 39: ...Operation LXT 380 Universal Transmitter Teledyne Analytical Instruments 20 Figure 3 1 LXT 380 Screen Map...
Страница 55: ...Operation LXT 380 Universal Transmitter Teledyne Analytical Instruments 36 Blank Page...
Страница 69: ...Maintenance LXT 380 Universal Transmitter Teledyne Analytical Instruments 50 Blank Page...
Страница 70: ...LXT 380 Universal Transmitter Troubleshooting Teledyne Analytical Instruments 51 6 0 TROUBLESHOOTING...
Страница 71: ...Troubleshooting LXT 380 Universal Transmitter Teledyne Analytical Instruments 52 Blank Page...
Страница 73: ...Parts and Accessories LXT 380 Universal Transmitter Teledyne Analytical Instruments 54 Blank Page...
Страница 78: ...LXT 380 Universal Transmitter SP3X Sensors Teledyne Analytical Instruments 59...
Страница 90: ...LXT 380 Universal Transmitter Appendix Teledyne Analytical Instruments 71 Appendix A Auto Cal Buffer Tables...
Страница 91: ...Appendix LXT 380 Universal Transmitter Teledyne Analytical Instruments 72 B LXT 380 Hart Menu...
Страница 92: ...LXT 380 Universal Transmitter Appendix Teledyne Analytical Instruments 73...
Страница 96: ...LXT 380 Universal Transmitter Appendix Teledyne Analytical Instruments 77...
Страница 97: ...Appendix LXT 380 Universal Transmitter Teledyne Analytical Instruments 78...
Страница 98: ...LXT 380 Universal Transmitter Appendix Teledyne Analytical Instruments 79 Fault Status WARNING STATUS...
Страница 99: ...Appendix LXT 380 Universal Transmitter Teledyne Analytical Instruments 80 Sensor Type...
Страница 101: ...Appendix LXT 380 Universal Transmitter Teledyne Analytical Instruments 82 E Drawings...
Страница 102: ...LXT 380 Universal Transmitter Appendix Teledyne Analytical Instruments 83...
Страница 103: ...Appendix LXT 380 Universal Transmitter Teledyne Analytical Instruments 84...