LXT-380 Universal Transmitter
Introduction
Teledyne Analytical Instruments
3
•
Do not install this equipment into a Zone 0 location.
•
Substitution of components is NOT PERMITTED and may
impact intrinsic safety.
•
Ensure a minimum 5 full threads of engagement.
•
Connect and verify green wire to LXT-380 grounding lug.
1.3 STOPPING PLUG 9310062
IECEx: SIR 07.0082X
ATEX: SIRA 07ATEX1240X
1180
II 2 G D/ I M2
Ex d I/IIC, Ex e I/II
Ex tD A21 IP6X
Stopping plug P/N 9310062 is intended for installation in
hazardous locations with Zone 1, Ex d Gb classification. The plug is
factory installed into the LXT-380 flame-proof housing to maintain the
flame-proof integrity for unused wiring ports in single channel
configuration. Do not remove or modify the plug or flame-path of the
plug. Verify that the plug is tightened to a torque of 55Nm (41ft-lbs).
•
Ensure the installation complies with all local, state and national
codes.
•
Refer to Control Drawing T1700004 for specific install
requirements. (See Appendix).
•
DO NOT substitute or alter this component as it may impact
flame-proof integrity.
1.4 SP3X SENSOR
IECEx: ETL 17.0030X
ATEX: ITS17ATEX101945X
1180
II 1 G Exia IIC T4 Ga
2
Содержание LXT-380
Страница 2: ......
Страница 4: ......
Страница 5: ......
Страница 6: ......
Страница 19: ...LXT 380 Universal Transmitter Teledyne Analytical Instruments xiv Blank Page...
Страница 27: ...Introduction LXT 380 Universal Transmitter Teledyne Analytical Instruments 8 1 8 APPROVALS 1 8 1 ENCLOSURE...
Страница 39: ...Operation LXT 380 Universal Transmitter Teledyne Analytical Instruments 20 Figure 3 1 LXT 380 Screen Map...
Страница 55: ...Operation LXT 380 Universal Transmitter Teledyne Analytical Instruments 36 Blank Page...
Страница 69: ...Maintenance LXT 380 Universal Transmitter Teledyne Analytical Instruments 50 Blank Page...
Страница 70: ...LXT 380 Universal Transmitter Troubleshooting Teledyne Analytical Instruments 51 6 0 TROUBLESHOOTING...
Страница 71: ...Troubleshooting LXT 380 Universal Transmitter Teledyne Analytical Instruments 52 Blank Page...
Страница 73: ...Parts and Accessories LXT 380 Universal Transmitter Teledyne Analytical Instruments 54 Blank Page...
Страница 78: ...LXT 380 Universal Transmitter SP3X Sensors Teledyne Analytical Instruments 59...
Страница 90: ...LXT 380 Universal Transmitter Appendix Teledyne Analytical Instruments 71 Appendix A Auto Cal Buffer Tables...
Страница 91: ...Appendix LXT 380 Universal Transmitter Teledyne Analytical Instruments 72 B LXT 380 Hart Menu...
Страница 92: ...LXT 380 Universal Transmitter Appendix Teledyne Analytical Instruments 73...
Страница 96: ...LXT 380 Universal Transmitter Appendix Teledyne Analytical Instruments 77...
Страница 97: ...Appendix LXT 380 Universal Transmitter Teledyne Analytical Instruments 78...
Страница 98: ...LXT 380 Universal Transmitter Appendix Teledyne Analytical Instruments 79 Fault Status WARNING STATUS...
Страница 99: ...Appendix LXT 380 Universal Transmitter Teledyne Analytical Instruments 80 Sensor Type...
Страница 101: ...Appendix LXT 380 Universal Transmitter Teledyne Analytical Instruments 82 E Drawings...
Страница 102: ...LXT 380 Universal Transmitter Appendix Teledyne Analytical Instruments 83...
Страница 103: ...Appendix LXT 380 Universal Transmitter Teledyne Analytical Instruments 84...