LXT-380 Universal Transmitter
Calibration
Teledyne Analytical Instruments
47
around 6000 μS (correlation factor 0.1667). All three solutions have a
TDS of 1000 ppm but the conductivities are 1400 μS, 2000 μS and 6000
μS. A TDS measurement is only valid for a solution with the same
chemical make up as the solution used for calibration.
Calibration is accomplished in two steps; Step 1 Conductivity
AUTO Calibration of the sensor (see section 4.7.1 above) and then Step
2 correlation of conductivity to the TDS. The second step can be done
by entering the correlation factor into the CONFIG →SENSOR→ TDS
menu or by standardizing the sensor in a solution of known TDS in the
CAL→STAND menu by entering the TDS value. Since the correlation
of conductivity to concentration is not linear it is best to calibrate the
sensor near the measured value.
4.6.1 STANDARDIZE
With the sensor in the process solution, take a grab sample from the
process and determine the TDS using a qualified laboratory method.
4.7 RESISTIVITY SENSORS
AUTO Calibration recognizes Air for zero point (Cal 1) and a Meg-
Ohm process solution for the span (Cal 2). Use the actual process
solution or a solution near the resistivity of the process solution. The
example below uses air and 15 MΩ. The Cal 1, air calibration value is
very stable and need not be done for every calibration.
Содержание LXT-380
Страница 2: ......
Страница 4: ......
Страница 5: ......
Страница 6: ......
Страница 19: ...LXT 380 Universal Transmitter Teledyne Analytical Instruments xiv Blank Page...
Страница 27: ...Introduction LXT 380 Universal Transmitter Teledyne Analytical Instruments 8 1 8 APPROVALS 1 8 1 ENCLOSURE...
Страница 39: ...Operation LXT 380 Universal Transmitter Teledyne Analytical Instruments 20 Figure 3 1 LXT 380 Screen Map...
Страница 55: ...Operation LXT 380 Universal Transmitter Teledyne Analytical Instruments 36 Blank Page...
Страница 69: ...Maintenance LXT 380 Universal Transmitter Teledyne Analytical Instruments 50 Blank Page...
Страница 70: ...LXT 380 Universal Transmitter Troubleshooting Teledyne Analytical Instruments 51 6 0 TROUBLESHOOTING...
Страница 71: ...Troubleshooting LXT 380 Universal Transmitter Teledyne Analytical Instruments 52 Blank Page...
Страница 73: ...Parts and Accessories LXT 380 Universal Transmitter Teledyne Analytical Instruments 54 Blank Page...
Страница 78: ...LXT 380 Universal Transmitter SP3X Sensors Teledyne Analytical Instruments 59...
Страница 90: ...LXT 380 Universal Transmitter Appendix Teledyne Analytical Instruments 71 Appendix A Auto Cal Buffer Tables...
Страница 91: ...Appendix LXT 380 Universal Transmitter Teledyne Analytical Instruments 72 B LXT 380 Hart Menu...
Страница 92: ...LXT 380 Universal Transmitter Appendix Teledyne Analytical Instruments 73...
Страница 96: ...LXT 380 Universal Transmitter Appendix Teledyne Analytical Instruments 77...
Страница 97: ...Appendix LXT 380 Universal Transmitter Teledyne Analytical Instruments 78...
Страница 98: ...LXT 380 Universal Transmitter Appendix Teledyne Analytical Instruments 79 Fault Status WARNING STATUS...
Страница 99: ...Appendix LXT 380 Universal Transmitter Teledyne Analytical Instruments 80 Sensor Type...
Страница 101: ...Appendix LXT 380 Universal Transmitter Teledyne Analytical Instruments 82 E Drawings...
Страница 102: ...LXT 380 Universal Transmitter Appendix Teledyne Analytical Instruments 83...
Страница 103: ...Appendix LXT 380 Universal Transmitter Teledyne Analytical Instruments 84...