LXT-380 Universal Transmitter
Calibration
Teledyne Analytical Instruments
37
4.0 CALIBRATION
The Model LXT-380 transmitter provides three methods of
calibration:
1.
Auto Calibration
2.
Standardize Calibration
3.
Manual Calibration
4.0.1 AUTO CALIBRATION DESCRIPTION
Auto calibration is the primary calibration method for all
measurements. AUTO calibration automatically recognizes the
calibration solution the sensor is in and proposes the actual temperature
compensated value for acceptance. AUTO calibration can be a single
point or two point calibration. A single point calibration sets the zero
point or offset value of the sensor. The second calibration sets the slope
or span of the sensor.
When the AUTO key is pressed the transmitter displays the PV
(Process Variable) and the associated mV signal from the sensor. When
the reading has stabilized a calibration value is
AUTOmatically proposed, i.e. 7.00 pH, 10 ppm
Fluoride ion, 0.00 mg/L Dissolved Oxygen.
The user is prompted to accept the proposed
calibration value or enter and accept another
value
. Once Cal 1 is accepted the user is ask to
continue to Cal 2, yes/no. If yes, then a second calibration value is
proposed when the sensor has stabilized in the second calibration
solution. Accept the value and the calibration is complete.
At the end of each calibration the Offset and Slope are displayed in
the respective units, pH, mV, ppm, mg/l.
Содержание LXT-380
Страница 2: ......
Страница 4: ......
Страница 5: ......
Страница 6: ......
Страница 19: ...LXT 380 Universal Transmitter Teledyne Analytical Instruments xiv Blank Page...
Страница 27: ...Introduction LXT 380 Universal Transmitter Teledyne Analytical Instruments 8 1 8 APPROVALS 1 8 1 ENCLOSURE...
Страница 39: ...Operation LXT 380 Universal Transmitter Teledyne Analytical Instruments 20 Figure 3 1 LXT 380 Screen Map...
Страница 55: ...Operation LXT 380 Universal Transmitter Teledyne Analytical Instruments 36 Blank Page...
Страница 69: ...Maintenance LXT 380 Universal Transmitter Teledyne Analytical Instruments 50 Blank Page...
Страница 70: ...LXT 380 Universal Transmitter Troubleshooting Teledyne Analytical Instruments 51 6 0 TROUBLESHOOTING...
Страница 71: ...Troubleshooting LXT 380 Universal Transmitter Teledyne Analytical Instruments 52 Blank Page...
Страница 73: ...Parts and Accessories LXT 380 Universal Transmitter Teledyne Analytical Instruments 54 Blank Page...
Страница 78: ...LXT 380 Universal Transmitter SP3X Sensors Teledyne Analytical Instruments 59...
Страница 90: ...LXT 380 Universal Transmitter Appendix Teledyne Analytical Instruments 71 Appendix A Auto Cal Buffer Tables...
Страница 91: ...Appendix LXT 380 Universal Transmitter Teledyne Analytical Instruments 72 B LXT 380 Hart Menu...
Страница 92: ...LXT 380 Universal Transmitter Appendix Teledyne Analytical Instruments 73...
Страница 96: ...LXT 380 Universal Transmitter Appendix Teledyne Analytical Instruments 77...
Страница 97: ...Appendix LXT 380 Universal Transmitter Teledyne Analytical Instruments 78...
Страница 98: ...LXT 380 Universal Transmitter Appendix Teledyne Analytical Instruments 79 Fault Status WARNING STATUS...
Страница 99: ...Appendix LXT 380 Universal Transmitter Teledyne Analytical Instruments 80 Sensor Type...
Страница 101: ...Appendix LXT 380 Universal Transmitter Teledyne Analytical Instruments 82 E Drawings...
Страница 102: ...LXT 380 Universal Transmitter Appendix Teledyne Analytical Instruments 83...
Страница 103: ...Appendix LXT 380 Universal Transmitter Teledyne Analytical Instruments 84...