2.2 WIRING
The LXT-380 installation consists of a flame-proof transmitter with
enclosure; an integral intrinsically-safe energy limiting barrier and
flame-proof seal; and an Intrinsically-safe sensor. Substitution of parts
or unauthorized repairs are prohibited and will invalidate certification.
Examples of unauthorized repairs include flame-path alteration or
modification of flame-proof components. Omission of components,
including but not limited to flame-proof conduit, flame-proof seals and
stopping plugs is not permitted. Flame-proof stopping plug shall be
tightened to a specified torque of 55Nm (41ft-lbs). Refer to local, state
and national codes for specific installation requirements and refer to
additional documents as necessary (NFPA33, NEC 500 and 516, IEC
60079-14).
Electrical wiring should only be conducted by qualified personnel.
See the LXT-380 Wiring Diagram in Figure 2.2.
Содержание LXT-380
Страница 2: ......
Страница 4: ......
Страница 5: ......
Страница 6: ......
Страница 19: ...LXT 380 Universal Transmitter Teledyne Analytical Instruments xiv Blank Page...
Страница 27: ...Introduction LXT 380 Universal Transmitter Teledyne Analytical Instruments 8 1 8 APPROVALS 1 8 1 ENCLOSURE...
Страница 39: ...Operation LXT 380 Universal Transmitter Teledyne Analytical Instruments 20 Figure 3 1 LXT 380 Screen Map...
Страница 55: ...Operation LXT 380 Universal Transmitter Teledyne Analytical Instruments 36 Blank Page...
Страница 69: ...Maintenance LXT 380 Universal Transmitter Teledyne Analytical Instruments 50 Blank Page...
Страница 70: ...LXT 380 Universal Transmitter Troubleshooting Teledyne Analytical Instruments 51 6 0 TROUBLESHOOTING...
Страница 71: ...Troubleshooting LXT 380 Universal Transmitter Teledyne Analytical Instruments 52 Blank Page...
Страница 73: ...Parts and Accessories LXT 380 Universal Transmitter Teledyne Analytical Instruments 54 Blank Page...
Страница 78: ...LXT 380 Universal Transmitter SP3X Sensors Teledyne Analytical Instruments 59...
Страница 90: ...LXT 380 Universal Transmitter Appendix Teledyne Analytical Instruments 71 Appendix A Auto Cal Buffer Tables...
Страница 91: ...Appendix LXT 380 Universal Transmitter Teledyne Analytical Instruments 72 B LXT 380 Hart Menu...
Страница 92: ...LXT 380 Universal Transmitter Appendix Teledyne Analytical Instruments 73...
Страница 96: ...LXT 380 Universal Transmitter Appendix Teledyne Analytical Instruments 77...
Страница 97: ...Appendix LXT 380 Universal Transmitter Teledyne Analytical Instruments 78...
Страница 98: ...LXT 380 Universal Transmitter Appendix Teledyne Analytical Instruments 79 Fault Status WARNING STATUS...
Страница 99: ...Appendix LXT 380 Universal Transmitter Teledyne Analytical Instruments 80 Sensor Type...
Страница 101: ...Appendix LXT 380 Universal Transmitter Teledyne Analytical Instruments 82 E Drawings...
Страница 102: ...LXT 380 Universal Transmitter Appendix Teledyne Analytical Instruments 83...
Страница 103: ...Appendix LXT 380 Universal Transmitter Teledyne Analytical Instruments 84...