LXT-380 Universal Transmitter
Operation
Teledyne Analytical Instruments
21
Double tap any Selection/Adjustment key to enter the HOME
Menu Screen. Five menu choices will appear,
CAL, CONFIG, INFO,
SIM and HOLD
. Each of the Menus is detailed below.
3.2.1 HOLD (OUTPUT HOLD)
Actuating the HOLD Key activates the HOLD function, HOLD is
ON, displayed.
•
Freezes the 4-20 mA output at the last
value prior to activation
•
Freezes optional Alarm Relays in the
current state
•
While in the HOLD mode the % Output
display toggles between the last value and
HOLD
Actuating HOLD again turns the hold function off, Hold is OFF,
displayed. The HOLD function remains ON until it is turned OFF. (See
Time Out in CONFIG>XMTR>OUTPUT>HOLD).
3.2.2 CAL (CALIBRATION MENU)
Four options are available,
AUTO, STAND, MANUAL and
TEMP.
On dual channel instruments choose Sensor 1 or Sensor 2 when
prompted.
The first screen asks, “
Is this a New Sensor,
YES / NO
”. If YES the calibration history from
the previous sensor is cleared from memory and a
new register is started, if NO then the calibration
is written to the memory stack, three sets of data
are stored.
•
AUTO
is a two point calibration. The calibration proceeds in
two steps, Auto Cal 1 is an offset calibration and Auto Cal 2 is a
slope calibration. Auto Cal provides automatic solution
recognition of the calibration solutions used for each
measurement in accordance with the following list:
1. pH Calibration Buffers (US Standard), pH 4.01, pH 7.00
and pH 10.00 (see Appendix A).
Содержание LXT-380
Страница 2: ......
Страница 4: ......
Страница 5: ......
Страница 6: ......
Страница 19: ...LXT 380 Universal Transmitter Teledyne Analytical Instruments xiv Blank Page...
Страница 27: ...Introduction LXT 380 Universal Transmitter Teledyne Analytical Instruments 8 1 8 APPROVALS 1 8 1 ENCLOSURE...
Страница 39: ...Operation LXT 380 Universal Transmitter Teledyne Analytical Instruments 20 Figure 3 1 LXT 380 Screen Map...
Страница 55: ...Operation LXT 380 Universal Transmitter Teledyne Analytical Instruments 36 Blank Page...
Страница 69: ...Maintenance LXT 380 Universal Transmitter Teledyne Analytical Instruments 50 Blank Page...
Страница 70: ...LXT 380 Universal Transmitter Troubleshooting Teledyne Analytical Instruments 51 6 0 TROUBLESHOOTING...
Страница 71: ...Troubleshooting LXT 380 Universal Transmitter Teledyne Analytical Instruments 52 Blank Page...
Страница 73: ...Parts and Accessories LXT 380 Universal Transmitter Teledyne Analytical Instruments 54 Blank Page...
Страница 78: ...LXT 380 Universal Transmitter SP3X Sensors Teledyne Analytical Instruments 59...
Страница 90: ...LXT 380 Universal Transmitter Appendix Teledyne Analytical Instruments 71 Appendix A Auto Cal Buffer Tables...
Страница 91: ...Appendix LXT 380 Universal Transmitter Teledyne Analytical Instruments 72 B LXT 380 Hart Menu...
Страница 92: ...LXT 380 Universal Transmitter Appendix Teledyne Analytical Instruments 73...
Страница 96: ...LXT 380 Universal Transmitter Appendix Teledyne Analytical Instruments 77...
Страница 97: ...Appendix LXT 380 Universal Transmitter Teledyne Analytical Instruments 78...
Страница 98: ...LXT 380 Universal Transmitter Appendix Teledyne Analytical Instruments 79 Fault Status WARNING STATUS...
Страница 99: ...Appendix LXT 380 Universal Transmitter Teledyne Analytical Instruments 80 Sensor Type...
Страница 101: ...Appendix LXT 380 Universal Transmitter Teledyne Analytical Instruments 82 E Drawings...
Страница 102: ...LXT 380 Universal Transmitter Appendix Teledyne Analytical Instruments 83...
Страница 103: ...Appendix LXT 380 Universal Transmitter Teledyne Analytical Instruments 84...