LXT-380 Universal Transmitter
SP3X Sensors
Teledyne Analytical Instruments
61
removing hard water scales and metallic deposits or spray washing for
flocculants and biofilms.
8.3.1 ELECTRODE CARTRIDGE INSTALLATION
Unless ordered separately, electrode cartridges are generally
shipped installed in a sensor. Sensors ordered without an electrode are
shipped with a shipping plug to keep contamination from getting inside
the sensor during shipment or storage. The following procedure explains
how to install the electrode cartridge in the sensor assembly:
1.
Remove the shipping plug by turning it counterclockwise.
2.
Remove the electrode cartridge from the protective soaker
boot.
Be careful not to flex the electrode body while removing
the tape and the protective boot.
3.
Rinse the electrode tip in tap water and wipe the electrode
body dry then lubricate the o-ring seals with the included
lubricant.
Save the protective soaker boot in the event the
electrode must be stored at a future time
.
4.
Carefully insert the electrode cartridge into the sensor
assembly by turning until
hand tight
. The first o-ring, closest
to the front of the electrode, will be slightly visible if held
horizontally.
Note: If excess force is required during electrode installation,
check for proper thread engagement or for an obstruction.
8.3.2 ELECTRODE CARTRIDGE REPLACEMENT
Periodic replacement of the electrode cartridge is required for pH,
ORP and Specific Ion sensors. The following procedure explains how to
replace the electrode cartridge in the sensor assembly:
1.
Remove the electrode cartridge from the front of the sensor
assembly by turning it counterclockwise.
2.
For installation procedure follow steps 2, 3, and 4 in section
8.3.1 electrode cartridge installation.
Содержание LXT-380
Страница 2: ......
Страница 4: ......
Страница 5: ......
Страница 6: ......
Страница 19: ...LXT 380 Universal Transmitter Teledyne Analytical Instruments xiv Blank Page...
Страница 27: ...Introduction LXT 380 Universal Transmitter Teledyne Analytical Instruments 8 1 8 APPROVALS 1 8 1 ENCLOSURE...
Страница 39: ...Operation LXT 380 Universal Transmitter Teledyne Analytical Instruments 20 Figure 3 1 LXT 380 Screen Map...
Страница 55: ...Operation LXT 380 Universal Transmitter Teledyne Analytical Instruments 36 Blank Page...
Страница 69: ...Maintenance LXT 380 Universal Transmitter Teledyne Analytical Instruments 50 Blank Page...
Страница 70: ...LXT 380 Universal Transmitter Troubleshooting Teledyne Analytical Instruments 51 6 0 TROUBLESHOOTING...
Страница 71: ...Troubleshooting LXT 380 Universal Transmitter Teledyne Analytical Instruments 52 Blank Page...
Страница 73: ...Parts and Accessories LXT 380 Universal Transmitter Teledyne Analytical Instruments 54 Blank Page...
Страница 78: ...LXT 380 Universal Transmitter SP3X Sensors Teledyne Analytical Instruments 59...
Страница 90: ...LXT 380 Universal Transmitter Appendix Teledyne Analytical Instruments 71 Appendix A Auto Cal Buffer Tables...
Страница 91: ...Appendix LXT 380 Universal Transmitter Teledyne Analytical Instruments 72 B LXT 380 Hart Menu...
Страница 92: ...LXT 380 Universal Transmitter Appendix Teledyne Analytical Instruments 73...
Страница 96: ...LXT 380 Universal Transmitter Appendix Teledyne Analytical Instruments 77...
Страница 97: ...Appendix LXT 380 Universal Transmitter Teledyne Analytical Instruments 78...
Страница 98: ...LXT 380 Universal Transmitter Appendix Teledyne Analytical Instruments 79 Fault Status WARNING STATUS...
Страница 99: ...Appendix LXT 380 Universal Transmitter Teledyne Analytical Instruments 80 Sensor Type...
Страница 101: ...Appendix LXT 380 Universal Transmitter Teledyne Analytical Instruments 82 E Drawings...
Страница 102: ...LXT 380 Universal Transmitter Appendix Teledyne Analytical Instruments 83...
Страница 103: ...Appendix LXT 380 Universal Transmitter Teledyne Analytical Instruments 84...