LXT-380 Universal Transmitter
Calibration
Teledyne Analytical Instruments
43
4.3.3 STANDARDIZE
Leave the sensor in the process solution, take a grab sample from
the process and determine the Ion concentration.
4.4 SP3X DISSOLVED OXYGEN CALIBRATION
PROCEDURES
The dissolved oxygen AUTO Cal acknowledges zero ppm, mg/l, %
SAT for CAL 1 and the temperature compensated value for atmospheric
oxygen, 8.25 ppm, mg/l at 25°C or 100 % SAT for CAL 2.
The zero point is set by placing the sensor into an oxygen free
solution and verifying the displayed value drops to a value below 1 mV.
The sensor will take a few minutes to equilibrate to the zero oxygen
solution but for the highest accuracy it is best to wait 15-20 minutes
before initiating a calibration. The typical sensor will burn down to 0.5 –
0.7 mV in an hour or so in a zero ppm solution. A zero ppm O
2
solution
can be made by adding approximately 5 grams of sodium sulfite to a
liter of distilled water or purging the sample with nitrogen gas.
The slope (CAL 2) is set by placing the sensor in air saturated
distilled water or water saturated (100% humidity) air. The easiest
method is to suspend the sensor vertically in beaker with a ½” of water
in the bottom slightly above the water.
A STANDardize calibration adjusts the CAL 2 value, resetting the
slope of the sensor, mV/ppm.
Содержание LXT-380
Страница 2: ......
Страница 4: ......
Страница 5: ......
Страница 6: ......
Страница 19: ...LXT 380 Universal Transmitter Teledyne Analytical Instruments xiv Blank Page...
Страница 27: ...Introduction LXT 380 Universal Transmitter Teledyne Analytical Instruments 8 1 8 APPROVALS 1 8 1 ENCLOSURE...
Страница 39: ...Operation LXT 380 Universal Transmitter Teledyne Analytical Instruments 20 Figure 3 1 LXT 380 Screen Map...
Страница 55: ...Operation LXT 380 Universal Transmitter Teledyne Analytical Instruments 36 Blank Page...
Страница 69: ...Maintenance LXT 380 Universal Transmitter Teledyne Analytical Instruments 50 Blank Page...
Страница 70: ...LXT 380 Universal Transmitter Troubleshooting Teledyne Analytical Instruments 51 6 0 TROUBLESHOOTING...
Страница 71: ...Troubleshooting LXT 380 Universal Transmitter Teledyne Analytical Instruments 52 Blank Page...
Страница 73: ...Parts and Accessories LXT 380 Universal Transmitter Teledyne Analytical Instruments 54 Blank Page...
Страница 78: ...LXT 380 Universal Transmitter SP3X Sensors Teledyne Analytical Instruments 59...
Страница 90: ...LXT 380 Universal Transmitter Appendix Teledyne Analytical Instruments 71 Appendix A Auto Cal Buffer Tables...
Страница 91: ...Appendix LXT 380 Universal Transmitter Teledyne Analytical Instruments 72 B LXT 380 Hart Menu...
Страница 92: ...LXT 380 Universal Transmitter Appendix Teledyne Analytical Instruments 73...
Страница 96: ...LXT 380 Universal Transmitter Appendix Teledyne Analytical Instruments 77...
Страница 97: ...Appendix LXT 380 Universal Transmitter Teledyne Analytical Instruments 78...
Страница 98: ...LXT 380 Universal Transmitter Appendix Teledyne Analytical Instruments 79 Fault Status WARNING STATUS...
Страница 99: ...Appendix LXT 380 Universal Transmitter Teledyne Analytical Instruments 80 Sensor Type...
Страница 101: ...Appendix LXT 380 Universal Transmitter Teledyne Analytical Instruments 82 E Drawings...
Страница 102: ...LXT 380 Universal Transmitter Appendix Teledyne Analytical Instruments 83...
Страница 103: ...Appendix LXT 380 Universal Transmitter Teledyne Analytical Instruments 84...