Operation
LXT-380 Universal Transmitter
Teledyne Analytical Instruments
28
PV connection (pH, ORP, Ion), Reference connection and Diagnostic
connection.
On a dual channel instrument an asterisk
*
will be displayed in the
sensor name block, SENSOR 2 *. The SENTINEL function will only be
visible in the Single Screen mode, not in the Dual Screen mode. Toggle
the BACK Key from Dual → Sensor 1 → Sensor 2 to see the SENTINEL
function.
Configure SENTINEL Function
The SENTINEL function is located in the Configure
Sensor menu:
CONFIG → SENSOR → SENSOR 1 or 2 →MORE
→ SNTL
There are three values displayed:
V
o
= SENTINEL Voltage offset, The diagnostic
voltage of a new electrode
V
e
= SENTINEL expiration Voltage, V
o
+ Range
RANGE = The shift in the diagnostic voltage required
to trigger diagnostic fault.
The default values of V
o
= 0.0 mV and the Range = 60
mV are good for most situations. The starting diagnostic
voltage is typically near 0 mV on a new electrode and if the
readings have drifted 60 mV apart (a full pH unit) then
electrode service is required.
3.3 START UP GUIDE
Install and wire the LXT-380 Transmitter as described in Sections
2.1 and 2.2 above.
Connect the sensor to the transmitter as described in Section 2.2
above.
Supply power to the Model LXT-380 transmitter.
Verify the proper measurement type is displayed, pH, ORP or Ion.
The sensor automatically uploads the measured parameter, the
Содержание LXT-380
Страница 2: ......
Страница 4: ......
Страница 5: ......
Страница 6: ......
Страница 19: ...LXT 380 Universal Transmitter Teledyne Analytical Instruments xiv Blank Page...
Страница 27: ...Introduction LXT 380 Universal Transmitter Teledyne Analytical Instruments 8 1 8 APPROVALS 1 8 1 ENCLOSURE...
Страница 39: ...Operation LXT 380 Universal Transmitter Teledyne Analytical Instruments 20 Figure 3 1 LXT 380 Screen Map...
Страница 55: ...Operation LXT 380 Universal Transmitter Teledyne Analytical Instruments 36 Blank Page...
Страница 69: ...Maintenance LXT 380 Universal Transmitter Teledyne Analytical Instruments 50 Blank Page...
Страница 70: ...LXT 380 Universal Transmitter Troubleshooting Teledyne Analytical Instruments 51 6 0 TROUBLESHOOTING...
Страница 71: ...Troubleshooting LXT 380 Universal Transmitter Teledyne Analytical Instruments 52 Blank Page...
Страница 73: ...Parts and Accessories LXT 380 Universal Transmitter Teledyne Analytical Instruments 54 Blank Page...
Страница 78: ...LXT 380 Universal Transmitter SP3X Sensors Teledyne Analytical Instruments 59...
Страница 90: ...LXT 380 Universal Transmitter Appendix Teledyne Analytical Instruments 71 Appendix A Auto Cal Buffer Tables...
Страница 91: ...Appendix LXT 380 Universal Transmitter Teledyne Analytical Instruments 72 B LXT 380 Hart Menu...
Страница 92: ...LXT 380 Universal Transmitter Appendix Teledyne Analytical Instruments 73...
Страница 96: ...LXT 380 Universal Transmitter Appendix Teledyne Analytical Instruments 77...
Страница 97: ...Appendix LXT 380 Universal Transmitter Teledyne Analytical Instruments 78...
Страница 98: ...LXT 380 Universal Transmitter Appendix Teledyne Analytical Instruments 79 Fault Status WARNING STATUS...
Страница 99: ...Appendix LXT 380 Universal Transmitter Teledyne Analytical Instruments 80 Sensor Type...
Страница 101: ...Appendix LXT 380 Universal Transmitter Teledyne Analytical Instruments 82 E Drawings...
Страница 102: ...LXT 380 Universal Transmitter Appendix Teledyne Analytical Instruments 83...
Страница 103: ...Appendix LXT 380 Universal Transmitter Teledyne Analytical Instruments 84...