How to Operate Your Instrument
Gray - Data in horizontal or vertical blanking intervals.
White - EAV and SAV packets; other reserved values such as the XYZ word.
Yellow - Data outside the nominal allowed values.
Red - Data with illegal values.
ANC Data Display
Pressing the
MAIN
button calls up the Main pop-up menu. From the Main pop-up menu, you can select the
ANC Display
display under the
Measure
heading. The ANC Data Display shows the raw (undecoded) data words in the selected ancillary
data packets. The ancillary data is shown only in hexadecimal format. Since the display is updated at approximately
two-second intervals, rapid data changes cannot be detected.
Use the large knob and the Select button to
enter the DID and/or SDID values for the type
of ancillary data packet to be monitored. You
can also touch the readouts on the screen to
select either DID or SDID for control using
the large knob. DID and SDID values have
the following characteristics:
DID and SDID Values range between
0x01 and 0xff.
For DID values in the range 0x00 -- 0x7F
(Type 2 packets), the SDID values are
used.
For DID values in the range 0x80 -- 0xFF
(Type 1 packets), there are no SDID
values used (the SDID soft key is grayed
out).