How to Operate Your Instrument
To Set Gain, Horizontal Magni
fi
cation and Sweep
Each tile maintains its own settings independent of the other tiles. These settings include Gain, Horizontal Magnification
and Sweep. For instance, when you switch a tile to a different measurement, the Gain, Horizontal Magnification, and
Sweep settings will be changed to what they were the last time the selected measurement was displayed in the tile. As the
procedure below indicates, Gain, Horizontal Magnification, and Sweep do not apply to all display types.
To set Gain:
1.
Select the tile whose gain you want to
adjust. (This will restore the menu for the
tile if it is not currently displayed.)
2.
Push the MAG/GAIN button. (If
MAG/GAIN is not available for the
tile you have selected, a message is
displayed to let you know that MAG/GAIN
is not available.)
3.
To set a fixed gain, touch the
V Gain
readout and use the large knob to
change the gain. You can also touch one
of the fixed gain soft keys.
4.
To set the VAR(iable) Gain, touch the
Var
Gain Enable
soft key so it is checked.
Set the gain that you want using the
large knob.
28