Monitoring the SDI Physical Layer
3.
Touch the
Vert Var Gain Enable
soft key
to enable variable gain adjustment, and
use the general-purpose knob to set the
waveform height to 10 major graticule
divisions.
4.
Use the
VERT
knob to position the top
and bottom of the waveform on graticule
lines.
5.
Press the
MAIN
front-panel button.
6.
Touch the
Eye
soft key.
7.
Touch the
Cursors
soft key. The
instrument displays the cursors
submenu.
8.
Select the
Show Timing
soft key to
enable display of the voltage cursors.
Timing cursor T1 is selected by default.
9.
Use the general-purpose knob to position
timing cursor T1 at the crossing of the
rising edge of the Eye waveform and the
graticule line two divisions above the
bottom of the waveform.
10.
Select timing cursor T2.
11.
Use the general-purpose knob to position
the timing cursor T2 at the crossing of
the rising edge of the Eye waveform and
the graticule line two divisions below the
top of the waveform.
Quick Start User Manual
95