Operating Instructions
2–24
CSA8000 & TDS8000 Instruments and Sampling Modules
Connector Care.
Never attach a cable to a sampling-module connector if the cable
has a worn or damaged connector because you may damage the sampling-module
connector. Use extra care when attaching or removing a cable from the connec-
tors. Turn only the nut, not the cable. When attaching a cable to a sampling-mod-
ule connector, align the connectors carefully before turning the nut. Use light
finger pressure to make this initial connection. Then tighten the nut lightly with a
torque wrench.
NOTE
. For best repeatability and to prolong connector life, use a torque wrench
and tighten the connection to the range of 79–112 N
@
cm (7–10 lb
@
in).
If the sampling-module connectors will receive heavy use, such as in a produc-
tion environment, you should install adapters (such as a Tektronix 015-0549-xx
for 3.5 mm connectors) on the sampling module to make connections to the
device under test.
Each channel has a SELECT channel button and a yellow channel light. The
button operates as follows:
H
If the yellow channel light is on, the channel is acquiring a waveform.
H
If you press the button and the channel is not currently being acquired (for
any channel or math waveform), then the instrument activates (turns on) the
channel.
H
If you press the button and the channel is currently active as a channel
waveform, then the instrument selects the channel waveform.
H
If the channel waveform is already selected when you press the channel
button, the instrument turns the channel off.
The TEKPROBE connector provides support for accessories requiring
TEKPROBE SMA support at levels 1 and 2. The connector provides power and
control to attached accessories, by the main instrument.
On modules with TDR capability, the red TDR ON light indicates whether the
step generator is sending out a step through the signal connector. The main
instrument turns this on or off.
Channel Selection
TEKPROBE Connector
TDR On Indicator
Содержание CSA8000 Series
Страница 4: ......
Страница 18: ...Service Safety Summary xiv CSA8000 TDS8000 Instruments and Sampling Modules ...
Страница 22: ...Preface xviii CSA8000 TDS8000 Instruments and Sampling Modules ...
Страница 23: ......
Страница 59: ...80C00 Modules Specifications 1 36 CSA8000 TDS8000 Instruments and Sampling Modules back ...
Страница 63: ...80A01 Module Specifications 1 40 CSA8000 TDS8000 Instruments and Sampling Modules ...
Страница 64: ......
Страница 74: ...Installation 2 10 CSA8000 TDS8000 Instruments and Sampling Modules ...
Страница 93: ......
Страница 103: ...Theory of Operation 3 10 CSA8000 TDS8000 Instruments and Sampling Modules ...
Страница 104: ......
Страница 106: ...Performance Verification Procedures 4 2 CSA8000 TDS8000 Instruments and Sampling Modules ...
Страница 122: ...Brief Procedures 4 18 CSA8000 TDS8000 Instruments and Sampling Modules ...
Страница 126: ...Performance Tests 4 22 CSA8000 TDS8000 Instruments and Sampling Modules ...
Страница 218: ...80A01 Trigger Preamplifier Module 4 114 CSA8000 TDS8000 Instruments and Sampling Modules ...
Страница 219: ......
Страница 231: ...Maintenance 6 6 CSA8000 TDS8000 Instruments and Sampling Modules ...
Страница 291: ...Removal and Installation Procedures 6 66 CSA8000 TDS8000 Instruments and Sampling Modules ...
Страница 307: ...Repackaging Instructions 6 82 CSA8000 TDS8000 Instruments and Sampling Modules ...
Страница 308: ......
Страница 313: ......
Страница 315: ...Electrical Parts List 8 2 CSA8000 TDS8000 Instruments and Sampling Modules ...
Страница 316: ......